Comparing Patellofemoral Kinematics Assessed With a Novel Muscle Actuator System and an Oxford Rig Using Noncadaveric Knees

Total knee replacement (TKR) failure, low patient satisfaction and high revision surgery rates may stem from insufficient preclinical testing. Conventional joint motion simulators for preclinical testing of TKR implants manipulate a knee joint in force, displacement, or simulated muscle control. However, a rig capable of using all three control modes has yet to be described in literature. This study aimed to validate a novel platform, the muscle actuator system (MAS), that can generate gravity-dependent, quadriceps-controlled squatting motions representative of an Oxford rig knee simulator and is mounted onto a force/displacement-control-capable joint motion simulator. Synthetic knee joint phantoms were created that comprised revision TKR implants and key extensor and flexor mechanism analogues, but no ligaments. The combined system implemented a constant force vector acting from simulated hip-to-ankle coordinates, effectively replicating gravity as observed in an Oxford rig. Quadriceps forces and patellofemoral joint kinematics were measured to assess the performance of the MAS and these tests showed high levels of repeatability and reproducibility. Forces and kinematics measured at a nominal patellar tendon length, and with patella alta and baja, were compared against those measured under the same conditions using a conventional Oxford rig, the Pennsylvania State Knee Simulator (PSKS). There was disagreement in absolute kinematics and muscle forces, but similar trends resulting from changing prosthesis design or patellar tendon length.

Authors

Institutions

Publication Details

Journal
Journal of Biomechanical Engineering
Published
2024-12-12
DOI
https://doi.org/10.1115/1.4067400
Citations
1
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing Patellofemoral Kinematics Assessed With a Novel Muscle Actuator System and an Oxford Rig Using Noncadaveric Knees

Ilya Borukhov, Ryan Willing, Stephen J. Piazza, Brent A. Lanting et al.
1 citations
Journal of Biomechanical Engineering
Total Knee Arthroplasty Outcomes
0.57
article

Comparing Patellofemoral Kinematics Assessed With a Novel Muscle Actuator System and an Oxford Rig Using Noncadaveric Knees

Ilya Borukhov, Ryan Willing, Stephen J. Piazza, Brent A. Lanting, Alexandre Galley, Samira Vakili
article en
1 citations

Abstract

Total knee replacement (TKR) failure, low patient satisfaction and high revision surgery rates may stem from insufficient preclinical testing. Conventional joint motion simulators for preclinical testing of TKR implants manipulate a knee joint in force, displacement, or simulated muscle control. However, a rig capable of using all three control modes has yet to be described in literature. This study aimed to validate a novel platform, the muscle actuator system (MAS), that can generate gravity-dependent, quadriceps-controlled squatting motions representative of an Oxford rig knee simulator and is mounted onto a force/displacement-control-capable joint motion simulator. Synthetic knee joint phantoms were created that comprised revision TKR implants and key extensor and flexor mechanism analogues, but no ligaments. The combined system implemented a constant force vector acting from simulated hip-to-ankle coordinates, effectively replicating gravity as observed in an Oxford rig. Quadriceps forces and patellofemoral joint kinematics were measured to assess the performance of the MAS and these tests showed high levels of repeatability and reproducibility. Forces and kinematics measured at a nominal patellar tendon length, and with patella alta and baja, were compared against those measured under the same conditions using a conventional Oxford rig, the Pennsylvania State Knee Simulator (PSKS). There was disagreement in absolute kinematics and muscle forces, but similar trends resulting from changing prosthesis design or patellar tendon length.

Journal of Biomechanical EngineeringVol. 147(3)
Western University (CA), Pennsylvania State University (US), Stryker (United States) (US)
Openalex Percentile: Top 26%
Total Knee Arthroplasty Outcomes
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.