Optimization of the design of a single-sided magnetic resonance magnet array for improved T<sub>2</sub> estimation accuracy

Abstract Single-sided magnetic resonance (SSMR) technology has a compact structure and great potential for use in mobile and nondestructive detection of large aqueous materials. As the detection depth increases, the magnetic field gradient generated by the magnetic resonance magnet placed on one side weakens, leading to a large distortion of T 2 , thereby significantly impeding accurate estimation of the pore size distribution within the detection target. In this study, to address this problem, the distortion in the region of interest (ROI) is significantly reduced by optimizing the size and arrangement of the SSMR magnet array. First, to evaluate the T 2 distortion within the ROI, we developed a T 2 distortion loss function based on α and θ , which are key parameters of the spin echo (SE) signal. We then combined the nonlinear constraints with the loss function to form a Lagrangian function, which was solved via the Newton iteration method to obtain the magnet array parameters. The magnet array is called the AT magnet array. The T 2 distortion was quantitatively evaluated by calculating the SE signal generated by the AT magnet array in the ROI. Furthermore, we compared the detection results obtained with the AT and the semi-annular (SA) magnet array. The results showed that within the 3 cm × 3 cm target range of a 10 cm depth, the AT magnet array had a smaller T 2 distortion of only 1.3%. After the ROI was expanded to 5 cm × 5 cm, the T 2 distortion of the SE signal measured by the AT magnet array was 6%, which met the SSMR detection accuracy requirements in this range. Therefore, the optimized design of the magnet array based on α and θ provides a theoretical basis for obtaining SE signals with low T 2 distortion at large depths and over large areas via SSMR.

Authors

Institutions

Publication Details

Journal
Measurement Science and Technology
Published
2024-12-12
DOI
https://doi.org/10.1088/1361-6501/ad9e16
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of the design of a single-sided magnetic resonance magnet array for improved T2 estimation accuracy

Chunpeng Ren, Yunzhi Wang, Chuandong Jiang, Q. Jane Wang et al.
Measurement Science and Technology
Advanced MRI Techniques and Applications
article

Optimization of the design of a single-sided magnetic resonance magnet array for improved T2 estimation accuracy

Chunpeng Ren, Yunzhi Wang, Chuandong Jiang, Q. Jane Wang, Ruixin Miao, Xiangqian Yu
article en

Abstract

Abstract Single-sided magnetic resonance (SSMR) technology has a compact structure and great potential for use in mobile and nondestructive detection of large aqueous materials. As the detection depth increases, the magnetic field gradient generated by the magnetic resonance magnet placed on one side weakens, leading to a large distortion of T 2 , thereby significantly impeding accurate estimation of the pore size distribution within the detection target. In this study, to address this problem, the distortion in the region of interest (ROI) is significantly reduced by optimizing the size and arrangement of the SSMR magnet array. First, to evaluate the T 2 distortion within the ROI, we developed a T 2 distortion loss function based on α and θ , which are key parameters of the spin echo (SE) signal. We then combined the nonlinear constraints with the loss function to form a Lagrangian function, which was solved via the Newton iteration method to obtain the magnet array parameters. The magnet array is called the AT magnet array. The T 2 distortion was quantitatively evaluated by calculating the SE signal generated by the AT magnet array in the ROI. Furthermore, we compared the detection results obtained with the AT and the semi-annular (SA) magnet array. The results showed that within the 3 cm × 3 cm target range of a 10 cm depth, the AT magnet array had a smaller T 2 distortion of only 1.3%. After the ROI was expanded to 5 cm × 5 cm, the T 2 distortion of the SE signal measured by the AT magnet array was 6%, which met the SSMR detection accuracy requirements in this range. Therefore, the optimized design of the magnet array based on α and θ provides a theoretical basis for obtaining SE signals with low T 2 distortion at large depths and over large areas via SSMR.

Measurement Science and TechnologyVol. 36(1)
Ministry of Education of the People's Republic of China (CN), Hohai University (CN), Jilin University (CN), Geophysical Survey (RU)
National Natural Science Foundation of China
Openalex Percentile: Top 61%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.