Low dose contrast enhancement of biodegradable low-density stents by an approach balancing radiopaque coatings and beam filtration
<i>Objective.</i>Biodegradable cardiovascular stents made of thin, low atomic number metals (e.g. Zn, Mg, Fe) are now approved for clinical use. However, poor contrast under x-ray imaging leads to longer surgical times, high patient exposure, and sometimes stent misplacement. This study aimed at enhancing the visibility of low-<i>Z</i>metal stents under x-ray imaging, by combining high-<i>Z</i>metal coatings and beam filtration.<i>Approach.</i>Photon energy spectra from W-anode x-ray beams operated at 80 and 120 kVp, were generated by the SpekCalc and BEAMnrc softwares. The contrast produced by Fe stent struts (50-10<i>μ</i>m W coatings), as well as dose and air kerma values (by BEAMnrc), were simulated. Several types of beam hardening filters (Sn: 0.1, 0.2 mm; Cu: 0.2, 0.7 mm) were also applied. Then, Fe foils (50<i>µ</i>m) with W coatings (2-3<i>µ</i>m-thick) were fabricated by magnetosputtering. These samples were x-ray visualized, for quantification of contrast between W-coated and uncoated Fe samples. Fe struts (50<i>µ</i>m) were also coated with W (3.8 ± 0.2<i>µ</i>m), and stent-like objects were x-ray visualized.<i>Main results.</i>Fe samples attenuate 6.4% (120 kVp) and 10.1% (80 kVp) spectra photons, and 25% and 34.5% for W-coated Fe samples (SpekCalc). BEAMnrc calculations revealed the highest contrast improvement in a 120 kVp beam (36.4%, and 38.5%) for W-coated and uncoated Fe samples with Sn (0.2 mm), and Cu + Sn (0.2 + 0.2 mm) filters. Experimentally, the highest contrasts between Fe and W-Fe foils, were obtained with 0.2 mm Sn (77 ± 7% contrast increase at 80 kV). The dose was also strongly reduced (70% and 75%, for 80 and 120 kVp beams). Finally, for 3D Fe stents visualized at 80 kVp, the highest CNR and CNRD values were achieved with 0.1 mm Sn (18.5 × and 20.1 mGy<sup>-1</sup>; compared to 15.0 × and 12.0 mGy<sup>-1</sup>in no-filter condition).<i>Significance.</i>The contrast of Fe-based stents in x-ray imaging is improved by addition of a thin layer of W and beam filtration with Sn. The precision and rapidity of biodegradable stents implantation would be improved thereby, as well as the dose to patients.
Authors
- Marc‐André Fortin (ORCID: https://orcid.org/0000-0002-8601-466X)
- Magdalena Bazalova‐Carter (ORCID: https://orcid.org/0000-0002-9365-2889)
- Diego Mantovani (ORCID: https://orcid.org/0000-0001-9672-895X)
- Théophraste Lescot (ORCID: https://orcid.org/0000-0003-3638-3907)
- Samira Ravanbakhsh (ORCID: https://orcid.org/0000-0001-5200-4685)
- Souheib Zekraoui
Institutions
- University of Victoria (CA)
- Centre hospitalier universitaire de Québec (CA)
- Laboratoire des Matériaux Avancés (FR)
- Centre hospitalier de l'Université Laval (CA)
- Université Laval (CA)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2024-12-12
- DOI
- https://doi.org/10.1088/1361-6560/ad9e7b
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Natural Sciences and Engineering Research Council of Canada