Low dose contrast enhancement of biodegradable low-density stents by an approach balancing radiopaque coatings and beam filtration

<i>Objective.</i>Biodegradable cardiovascular stents made of thin, low atomic number metals (e.g. Zn, Mg, Fe) are now approved for clinical use. However, poor contrast under x-ray imaging leads to longer surgical times, high patient exposure, and sometimes stent misplacement. This study aimed at enhancing the visibility of low-<i>Z</i>metal stents under x-ray imaging, by combining high-<i>Z</i>metal coatings and beam filtration.<i>Approach.</i>Photon energy spectra from W-anode x-ray beams operated at 80 and 120 kVp, were generated by the SpekCalc and BEAMnrc softwares. The contrast produced by Fe stent struts (50-10<i>μ</i>m W coatings), as well as dose and air kerma values (by BEAMnrc), were simulated. Several types of beam hardening filters (Sn: 0.1, 0.2 mm; Cu: 0.2, 0.7 mm) were also applied. Then, Fe foils (50<i>µ</i>m) with W coatings (2-3<i>µ</i>m-thick) were fabricated by magnetosputtering. These samples were x-ray visualized, for quantification of contrast between W-coated and uncoated Fe samples. Fe struts (50<i>µ</i>m) were also coated with W (3.8 ± 0.2<i>µ</i>m), and stent-like objects were x-ray visualized.<i>Main results.</i>Fe samples attenuate 6.4% (120 kVp) and 10.1% (80 kVp) spectra photons, and 25% and 34.5% for W-coated Fe samples (SpekCalc). BEAMnrc calculations revealed the highest contrast improvement in a 120 kVp beam (36.4%, and 38.5%) for W-coated and uncoated Fe samples with Sn (0.2 mm), and Cu + Sn (0.2 + 0.2 mm) filters. Experimentally, the highest contrasts between Fe and W-Fe foils, were obtained with 0.2 mm Sn (77 ± 7% contrast increase at 80 kV). The dose was also strongly reduced (70% and 75%, for 80 and 120 kVp beams). Finally, for 3D Fe stents visualized at 80 kVp, the highest CNR and CNRD values were achieved with 0.1 mm Sn (18.5 × and 20.1 mGy<sup>-1</sup>; compared to 15.0 × and 12.0 mGy<sup>-1</sup>in no-filter condition).<i>Significance.</i>The contrast of Fe-based stents in x-ray imaging is improved by addition of a thin layer of W and beam filtration with Sn. The precision and rapidity of biodegradable stents implantation would be improved thereby, as well as the dose to patients.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-12-12
DOI
https://doi.org/10.1088/1361-6560/ad9e7b
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Low dose contrast enhancement of biodegradable low-density stents by an approach balancing radiopaque coatings and beam filtration

Marc‐André Fortin, Magdalena Bazalova‐Carter, Diego Mantovani, Théophraste Lescot et al.
1 links
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Low dose contrast enhancement of biodegradable low-density stents by an approach balancing radiopaque coatings and beam filtration

Marc‐André Fortin, Magdalena Bazalova‐Carter, Diego Mantovani, Théophraste Lescot, Samira Ravanbakhsh, Souheib Zekraoui
article en

Abstract

<i>Objective.</i>Biodegradable cardiovascular stents made of thin, low atomic number metals (e.g. Zn, Mg, Fe) are now approved for clinical use. However, poor contrast under x-ray imaging leads to longer surgical times, high patient exposure, and sometimes stent misplacement. This study aimed at enhancing the visibility of low-<i>Z</i>metal stents under x-ray imaging, by combining high-<i>Z</i>metal coatings and beam filtration.<i>Approach.</i>Photon energy spectra from W-anode x-ray beams operated at 80 and 120 kVp, were generated by the SpekCalc and BEAMnrc softwares. The contrast produced by Fe stent struts (50-10<i>μ</i>m W coatings), as well as dose and air kerma values (by BEAMnrc), were simulated. Several types of beam hardening filters (Sn: 0.1, 0.2 mm; Cu: 0.2, 0.7 mm) were also applied. Then, Fe foils (50<i>µ</i>m) with W coatings (2-3<i>µ</i>m-thick) were fabricated by magnetosputtering. These samples were x-ray visualized, for quantification of contrast between W-coated and uncoated Fe samples. Fe struts (50<i>µ</i>m) were also coated with W (3.8 ± 0.2<i>µ</i>m), and stent-like objects were x-ray visualized.<i>Main results.</i>Fe samples attenuate 6.4% (120 kVp) and 10.1% (80 kVp) spectra photons, and 25% and 34.5% for W-coated Fe samples (SpekCalc). BEAMnrc calculations revealed the highest contrast improvement in a 120 kVp beam (36.4%, and 38.5%) for W-coated and uncoated Fe samples with Sn (0.2 mm), and Cu + Sn (0.2 + 0.2 mm) filters. Experimentally, the highest contrasts between Fe and W-Fe foils, were obtained with 0.2 mm Sn (77 ± 7% contrast increase at 80 kV). The dose was also strongly reduced (70% and 75%, for 80 and 120 kVp beams). Finally, for 3D Fe stents visualized at 80 kVp, the highest CNR and CNRD values were achieved with 0.1 mm Sn (18.5 × and 20.1 mGy<sup>-1</sup>; compared to 15.0 × and 12.0 mGy<sup>-1</sup>in no-filter condition).<i>Significance.</i>The contrast of Fe-based stents in x-ray imaging is improved by addition of a thin layer of W and beam filtration with Sn. The precision and rapidity of biodegradable stents implantation would be improved thereby, as well as the dose to patients.

Physics in Medicine and BiologyVol. 70(2)
University of Victoria (CA), Centre hospitalier universitaire de Québec (CA), Laboratoire des Matériaux Avancés (FR), Centre hospitalier de l'Université Laval (CA), Université Laval (CA)
Natural Sciences and Engineering Research Council of Canada
Affordable and clean energy
Openalex Percentile: Top 79%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.