METAL DEBRIS RELEASE IN TOTAL KNEE ARTHROPLASTY: AN OVERLOOKED CLINICAL CONCERN

Objectives Several studies have reported elevated blood cobalt (Co) and chromium (Cr) concentrations in patients with total knee replacements (TKRs). Up to 44% of tissue samples taken from patients with failed TKRs exhibit histological evidence of metal sensitivity/ALVAL. In simulated conditions, metal particles contribute approximately 12% of total wear debris in TKR. We carried out this investigation to determine the source and quantity of metal release in TKRs. Design and Methods We analysed 225 explanted fixed-bearing TKRs (Attune, Genesis II, NexGen, PFC, and Vanguard) revised for any indication. These were analysed using peer-reviewed [coordinate measuring machine (CMM)] methodology to measure the volumetric wear of the polyethylene (PE) bearing surfaces and trays. The trays were analysed using 2D profilometry (surface roughness-Ra) and light microscopy. Histological and blood metal ion concentration analyses were performed in a sub-sample of patients. Results The median (IQR) PE wear rate was 14 (6 to 20) mm 3 /year. Microscopic examination of the superior surface of trays exhibited pitting on 132 (59%) of trays. There was a statistically significant (p<0.05) increase in Rvk on the pitted area of trays for each design, indicating material removal from the pits compared to the unpitted area. The inferior surface of 116(51%) of trays displayed polishing, indicative of abrasive wear. The median(range) Co and Cr concentrations were 2.5µg/l (0.2–69.4) and 1.7µg/l (0.5-12.5) respectively in 40 patients. Of the tissue samples examined in 30 patients, 6 had at least “mild”-ALVAL infiltrate. All corresponding “ALVAL” explants were found to be pitted and/or show evidence of loosening of the tray. Conclusion This study provides further evidence that CoCr release in TKR appears to be an under-appreciated cause of adverse clinical outcomes. The generation of metal particles was predominantly from the metal tray, which may explain elevated metal ions after TKRs, despite no direct metal-on-metal contact.

Authors

Publication Details

Journal
Orthopaedic Proceedings
Published
2024-12-12
DOI
https://doi.org/10.1302/1358-992x.2024.20.012
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

METAL DEBRIS RELEASE IN TOTAL KNEE ARTHROPLASTY: AN OVERLOOKED CLINICAL CONCERN

Stephen R. Wells, David Langton, Rohan Bhalekar, Matthew Nargol et al.
Orthopaedic Proceedings
Orthopaedic implants and arthroplasty
article

METAL DEBRIS RELEASE IN TOTAL KNEE ARTHROPLASTY: AN OVERLOOKED CLINICAL CONCERN

Stephen R. Wells, David Langton, Rohan Bhalekar, Matthew Nargol, Moreica Pabbruwe, Sonali Natu, Raghavendra Sidaginamale, A. V. F. Nargol, Susan Waller
article en

Abstract

Objectives Several studies have reported elevated blood cobalt (Co) and chromium (Cr) concentrations in patients with total knee replacements (TKRs). Up to 44% of tissue samples taken from patients with failed TKRs exhibit histological evidence of metal sensitivity/ALVAL. In simulated conditions, metal particles contribute approximately 12% of total wear debris in TKR. We carried out this investigation to determine the source and quantity of metal release in TKRs. Design and Methods We analysed 225 explanted fixed-bearing TKRs (Attune, Genesis II, NexGen, PFC, and Vanguard) revised for any indication. These were analysed using peer-reviewed [coordinate measuring machine (CMM)] methodology to measure the volumetric wear of the polyethylene (PE) bearing surfaces and trays. The trays were analysed using 2D profilometry (surface roughness-Ra) and light microscopy. Histological and blood metal ion concentration analyses were performed in a sub-sample of patients. Results The median (IQR) PE wear rate was 14 (6 to 20) mm 3 /year. Microscopic examination of the superior surface of trays exhibited pitting on 132 (59%) of trays. There was a statistically significant (p<0.05) increase in Rvk on the pitted area of trays for each design, indicating material removal from the pits compared to the unpitted area. The inferior surface of 116(51%) of trays displayed polishing, indicative of abrasive wear. The median(range) Co and Cr concentrations were 2.5µg/l (0.2–69.4) and 1.7µg/l (0.5-12.5) respectively in 40 patients. Of the tissue samples examined in 30 patients, 6 had at least “mild”-ALVAL infiltrate. All corresponding “ALVAL” explants were found to be pitted and/or show evidence of loosening of the tray. Conclusion This study provides further evidence that CoCr release in TKR appears to be an under-appreciated cause of adverse clinical outcomes. The generation of metal particles was predominantly from the metal tray, which may explain elevated metal ions after TKRs, despite no direct metal-on-metal contact.

Orthopaedic ProceedingsVol. 106-B(SUPP_20)
Openalex Percentile: Top 57%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.