Systematic Approach to Identifying Sources of Variation in CT Radiomics: A Phantom Study

Variations in CT parameters affect extracted radiomic features, potentially confounding disease phenotype analysis in multicenter studies. However, the effects of such variations are not yet clearly understood. The purpose of this study was to assess CT radiomic feature robustness across various acquisition parameters using both a large water phantom and a liver phantom with 3D-printed textures. In this study, scans of both phantoms were acquired on Siemens CT with varied slice thicknesses, doses, voltages, and reconstruction methods. By varying these imaging parameters, 36 distinct images were acquired of the phantom. ROIs of the texture inserts were manually segmented and, from these, 107 features belonging to 6 subgroups of texture methods were extracted using Pyradiomics. A heat map was constructed to identify robust metrics. Across both phantoms, slice thickness showed the strongest impact on radiomic features (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf {\beta}=\mathbf{-0.18}$</tex>, IQR: -0.36 to 0.33; 0.6mm reference), followed by reconstruction algorithm (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf{\beta}={-} {0.14}$</tex>, IQR: -0.36 to 0.31; FBP reference), and dose (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf{\beta}={-} {0.10}$</tex>, IQR: -0.23 to 0.20; 40% reduction reference). Texture-based methods like GLCM, GLDM, GLRLM, and GLSZM demonstrated greater robustness compared to first-order measures such as SD, CV, and IQR. Our results may serve as a guide for clinicians and researchers alike to help them select radiomic metrics that are largely immune to these variations in image acquisition parameters and reconstruction methods. Radiomics-based phantom studies help identify stable radiomic features, essential for distinguishing genuine diagnoses from false positives caused by imaging parameter variations in clinical settings.

Authors

Institutions

Publication Details

Published
2024-11-13
DOI
https://doi.org/10.1109/sipaim62974.2024.10783586
Citations
3
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.55
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Systematic Approach to Identifying Sources of Variation in CT Radiomics: A Phantom Study

David J. Goodenough, Steven Cen, Kristin Jensen, Bino Varghese et al.
3 citations
Advanced X-ray and CT Imaging
0.55
article

Systematic Approach to Identifying Sources of Variation in CT Radiomics: A Phantom Study

David J. Goodenough, Steven Cen, Kristin Jensen, Bino Varghese, Joshua Levy, Neha Yadav, Darryl Hwang, Xiaomeng Lei, Huawei Han, Andrew Turangan
article en
3 citations

Abstract

Variations in CT parameters affect extracted radiomic features, potentially confounding disease phenotype analysis in multicenter studies. However, the effects of such variations are not yet clearly understood. The purpose of this study was to assess CT radiomic feature robustness across various acquisition parameters using both a large water phantom and a liver phantom with 3D-printed textures. In this study, scans of both phantoms were acquired on Siemens CT with varied slice thicknesses, doses, voltages, and reconstruction methods. By varying these imaging parameters, 36 distinct images were acquired of the phantom. ROIs of the texture inserts were manually segmented and, from these, 107 features belonging to 6 subgroups of texture methods were extracted using Pyradiomics. A heat map was constructed to identify robust metrics. Across both phantoms, slice thickness showed the strongest impact on radiomic features (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf {\beta}=\mathbf{-0.18}$</tex>, IQR: -0.36 to 0.33; 0.6mm reference), followed by reconstruction algorithm (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf{\beta}={-} {0.14}$</tex>, IQR: -0.36 to 0.31; FBP reference), and dose (median <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mathbf{\beta}={-} {0.10}$</tex>, IQR: -0.23 to 0.20; 40% reduction reference). Texture-based methods like GLCM, GLDM, GLRLM, and GLSZM demonstrated greater robustness compared to first-order measures such as SD, CV, and IQR. Our results may serve as a guide for clinicians and researchers alike to help them select radiomic metrics that are largely immune to these variations in image acquisition parameters and reconstruction methods. Radiomics-based phantom studies help identify stable radiomic features, essential for distinguishing genuine diagnoses from false positives caused by imaging parameter variations in clinical settings.

University of Southern California (US), Oslo University Hospital (NO)
Openalex Percentile: Top 36%
Advanced X-ray and CT Imaging
0.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.