Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database
Our analysis of the MAUDE database showed that device delivery issues, bleeding, and vascular complications accounted for a smaller proportion of reports with Impella RP Flex than with its predecessor. However, device-related thrombosis and pump malfunction accounted for a higher proportion of reports with RP Flex. These data do not reflect the overall incidence of complications, but instead highlight the need for physicians and other healthcare providers to return devices to the manufacturer for inspection when a failure occurs.