Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database

Our analysis of the MAUDE database showed that device delivery issues, bleeding, and vascular complications accounted for a smaller proportion of reports with Impella RP Flex than with its predecessor. However, device-related thrombosis and pump malfunction accounted for a higher proportion of reports with RP Flex. These data do not reflect the overall incidence of complications, but instead highlight the need for physicians and other healthcare providers to return devices to the manufacturer for inspection when a failure occurs.

Authors

Institutions

Publication Details

Journal
Catheterization and Cardiovascular Interventions
Published
2024-12-11
DOI
https://doi.org/10.1002/ccd.31333
Citations
2
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
1.10
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database

Ron Waksman, Marta Lorente‐Ros, Ilan Merdler, Abhishek Chaturvedi et al.
2 citations
Catheterization and Cardiovascular Interventions
Venous Thromboembolism Diagnosis and Management
1.10
article

Adverse Events and Failure Modes Related to Impella RP/RP Flex: Insights From FDA MAUDE Database

Ron Waksman, Marta Lorente‐Ros, Ilan Merdler, Abhishek Chaturvedi, Kalyan R. Chitturi, Toby Rogers, Hayder Hashim, Brian C. Case, Ryan Wallace, Itsik Ben‐Dor
article en
2 citations

Abstract

Our analysis of the MAUDE database showed that device delivery issues, bleeding, and vascular complications accounted for a smaller proportion of reports with Impella RP Flex than with its predecessor. However, device-related thrombosis and pump malfunction accounted for a higher proportion of reports with RP Flex. These data do not reflect the overall incidence of complications, but instead highlight the need for physicians and other healthcare providers to return devices to the manufacturer for inspection when a failure occurs.

Catheterization and Cardiovascular InterventionsVol. 105(2)
MedStar Washington Hospital Center (US)
Good health and well-being
Openalex Percentile: Top 21%
Venous Thromboembolism Diagnosis and Management
1.10
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.