The impact of the xSPECT reconstruction algorithms on the recovery coefficients value for small tumors: a phantom study with 177Lu

Quantifying small tumors is still a challenge due to the partial volume effect (PVE). Although iterative reconstruction had promising results with a better recovery coefficient (RC), it suffers from the PVE. RC values typically depend on the reconstruction method, which may affect on <sup>177</sup>Lu quantifying. In this study, we investigated the effect of different reconstruction methods on RC values for <sup>177</sup>Lu SPECT/CT images.A water-filled cylindrical polymethylmethacrylate (PMMA) phantom without background radioactivity with two solid cylinders which represent the bone and soft tissue equivalent was used for SPECT/CT imaging. We placed syringes with different volumes containing <sup>177</sup>Lu in the vicinity of the cylinders as tumors. The iterative reconstructions of xSPECT, xSPECT Bone, and Flash3D was performed on SPECT/CT images.The effect of the reconstruction algorithms on the RC values was determined by calculating the radioactivity concentration in volumes of interest which were drawn manually and by a threshold method.The xSPECT Bone reconstruction has shown an RC value of approximately 100%, while the other reconstructions typically show an RC value of 93%. The RC values for all the reconstructions were decreased for smaller tumors. The smallest tumor showed a minimum RC between 52% and 72% compared to the largest tumor.In this study, it was possible to obtain an RC value close to 100% when the xSPECT Bone reconstruction algorithm was used. The xSPECT results were slightly lower RC values compared to the Flash3D algorithm. In using standard RC curves for tumors close to the bone, special care has to be taken since RC values relative to bone are higher than in tissue.

Authors

Institutions

Publication Details

Journal
Nuklearmedizin - NuclearMedicine
Published
2024-12-11
DOI
https://doi.org/10.1055/a-2418-2486
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of the xSPECT reconstruction algorithms on the recovery coefficients value for small tumors: a phantom study with 177Lu

Behrooz H. Yousefi, Robin de Nijs, Markus Luster, Parham Geramifar et al.
1 citations
Nuklearmedizin - NuclearMedicine
Medical Imaging Techniques and Applications
0.40
article

The impact of the xSPECT reconstruction algorithms on the recovery coefficients value for small tumors: a phantom study with 177Lu

Behrooz H. Yousefi, Robin de Nijs, Markus Luster, Parham Geramifar, Ali Ebrahimifard, Hossein Rajabi
article en
1 citations

Abstract

Quantifying small tumors is still a challenge due to the partial volume effect (PVE). Although iterative reconstruction had promising results with a better recovery coefficient (RC), it suffers from the PVE. RC values typically depend on the reconstruction method, which may affect on <sup>177</sup>Lu quantifying. In this study, we investigated the effect of different reconstruction methods on RC values for <sup>177</sup>Lu SPECT/CT images.A water-filled cylindrical polymethylmethacrylate (PMMA) phantom without background radioactivity with two solid cylinders which represent the bone and soft tissue equivalent was used for SPECT/CT imaging. We placed syringes with different volumes containing <sup>177</sup>Lu in the vicinity of the cylinders as tumors. The iterative reconstructions of xSPECT, xSPECT Bone, and Flash3D was performed on SPECT/CT images.The effect of the reconstruction algorithms on the RC values was determined by calculating the radioactivity concentration in volumes of interest which were drawn manually and by a threshold method.The xSPECT Bone reconstruction has shown an RC value of approximately 100%, while the other reconstructions typically show an RC value of 93%. The RC values for all the reconstructions were decreased for smaller tumors. The smallest tumor showed a minimum RC between 52% and 72% compared to the largest tumor.In this study, it was possible to obtain an RC value close to 100% when the xSPECT Bone reconstruction algorithm was used. The xSPECT results were slightly lower RC values compared to the Flash3D algorithm. In using standard RC curves for tumors close to the bone, special care has to be taken since RC values relative to bone are higher than in tissue.

Nuklearmedizin - NuclearMedicineVol. 64(02)
Tarbiat Modares University (IR), Philipps University of Marburg (DE), Copenhagen University Hospital (DK), Rigshospitalet (DK), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 32%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.