Histological assessment of thermal damage in porcine muscle induced by monopolar electrosurgical cutting devices during manual and robotic testing

Surgical cutting with electrosurgical tools facilitates tissue dissection and vessel sealing, preventing blood loss. The extent of tissue necrosis due to temperature elevations is dependent on the cutting technique, device design, coating properties and power settings, but the influence of these parameters is not fully understood. Here we conduct a comprehensive comparative analysis of thermal damage comparing (1) manual user-controlled and robotic electrosurgical cutting approaches for (2) varying electrodes and coatings, and power settings. We demonstrate that ceramic coating significantly enhanced cutting performance and cut quality and reduced lateral thermal damage, by 86.15% at 35 W and 65% at 50 W respectively. We provide quantitative assessment of the influence of surgical variability on thermal damage, comparing robotic and manual electrosurgical cutting. Robotic cutting with one ceramic electrosurgical coated device reduced thermal damage (midline - 47.42%, lateral - 33.06%), whereas for the other coated electrode the thermal spread increased (midline - 66.57%, lateral -245.72). Thus, thermal damage performance was strongly influenced by surgical variability and the specific characteristics of each device. Together, these results provide an enhanced understanding of potential mechanisms determining electrosurgical outcomes. Understanding of these interdependencies and mechanisms of action linked to a specific electrosurgical system is essential for successful tissue resection.

Authors

Institutions

Publication Details

Journal
International Journal of Hyperthermia
Published
2024-12-11
DOI
https://doi.org/10.1080/02656736.2024.2439549
Citations
1
Primary Topic
Additive Manufacturing Materials and Processes
Type
article
Field-Weighted Citation Impact
0.21

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Histological assessment of thermal damage in porcine muscle induced by monopolar electrosurgical cutting devices during manual and robotic testing

Laoise M. McNamara, Vincent J. Casey, Elzbieta Ewertowska, Robert Whiting et al.
1 citations
International Journal of Hyperthermia
Additive Manufacturing Materials and Processes
0.21
article

Histological assessment of thermal damage in porcine muscle induced by monopolar electrosurgical cutting devices during manual and robotic testing

Laoise M. McNamara, Vincent J. Casey, Elzbieta Ewertowska, Robert Whiting, Micheal Burke, Laura Frey, Paul Sheridan, Ben Row, Bryan Deeny
article en
1 citations

Abstract

Surgical cutting with electrosurgical tools facilitates tissue dissection and vessel sealing, preventing blood loss. The extent of tissue necrosis due to temperature elevations is dependent on the cutting technique, device design, coating properties and power settings, but the influence of these parameters is not fully understood. Here we conduct a comprehensive comparative analysis of thermal damage comparing (1) manual user-controlled and robotic electrosurgical cutting approaches for (2) varying electrodes and coatings, and power settings. We demonstrate that ceramic coating significantly enhanced cutting performance and cut quality and reduced lateral thermal damage, by 86.15% at 35 W and 65% at 50 W respectively. We provide quantitative assessment of the influence of surgical variability on thermal damage, comparing robotic and manual electrosurgical cutting. Robotic cutting with one ceramic electrosurgical coated device reduced thermal damage (midline - 47.42%, lateral - 33.06%), whereas for the other coated electrode the thermal spread increased (midline - 66.57%, lateral -245.72). Thus, thermal damage performance was strongly influenced by surgical variability and the specific characteristics of each device. Together, these results provide an enhanced understanding of potential mechanisms determining electrosurgical outcomes. Understanding of these interdependencies and mechanisms of action linked to a specific electrosurgical system is essential for successful tissue resection.

International Journal of HyperthermiaVol. 41(1)
Ollscoil na Gaillimhe – University of Galway (IE), Stryker (Ireland) (IE)
Science Foundation Ireland, European Regional Development Fund
Openalex Percentile: Top 49%
Additive Manufacturing Materials and Processes
0.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.