Miniature Mass Spectrometry for Point-of-Care Testing the GPIIb/IIIa Inhibitor Tirofiban during Perioperative Period of Percutaneous Coronary Intervention

Precise administration of tirofiban must be carefully considered to achieve the best treatment efficacy and maximize safety for patients. Herein, a paper spray ionization (PSI) linear ion trap (LIT) portable mass spectrometer (pMS) based point-of-care testing (POCT) technique was developed for on-site sampling, clinical testing, and immediate analysis of tirofiban blood drug concentrations. The results showed that tirofiban formed a significant and stable parent ion peak at <i>m</i>/<i>z</i> 441.3 by MS1 full scan in positive ion mode, which fragmented into product ions at <i>m</i>/<i>z</i> 395.4, <i>m</i>/<i>z</i> 321.3, <i>m</i>/<i>z</i> 276.3, and <i>m</i>/<i>z</i> 260.3 through collision-induced dissociation (CID) in MS/MS. To improve ion response, the parameters were optimized to the voltages of ionization 3600 V, ion isolation 1 (ISO1) 8 V, ion isolation 2 (ISO2) 2 V, and collision-induced dissociation (CID) 3 V. This method was validated, and the limit of detection (LOD) and limit of quantification (LOQ) were found to be 10.1 and 33.7 μg·L<sup>-1</sup>, respectively. For precision, it had the relative standard deviation (RSD) of interday precision of 4.8 to 6.7% and the RSD of intraday precision of 7.8 to 8.3%. The recovery of the method ranged from 87.5 to 93.4%. Although matrix effects in blood samples had some inhibitory effects on the target signal formation, the method compensated for part of the matrix effects by establishing a matrix-matched calibration curve, which exhibited good linearity with a <i>R</i>2 of 0.9987. Finally, the method was applied to the detection of tirofiban in clinically collected blood samples. Out of 12 samples, ten had tirofiban concentrations between 35.4 and 72.1 μg·L<sup>-1</sup> while the remaining two were below the LOQ. The method needs further optimization and validation in the future to improve its sensitivity and stability.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2024-12-10
DOI
https://doi.org/10.1021/acsomega.4c06581
Citations
4
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
1.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Miniature Mass Spectrometry for Point-of-Care Testing the GPIIb/IIIa Inhibitor Tirofiban during Perioperative Period of Percutaneous Coronary Intervention

Dongchen Zhou, Huizhi Ding, Weizong Zhang, Haixing Wang et al.
4 citations
ACS Omega
Acute Myocardial Infarction Research
1.61
article

Miniature Mass Spectrometry for Point-of-Care Testing the GPIIb/IIIa Inhibitor Tirofiban during Perioperative Period of Percutaneous Coronary Intervention

Dongchen Zhou, Huizhi Ding, Weizong Zhang, Haixing Wang, Yong Liu, Shiqi Wang, Yunfeng Zhang, Qing Shen, Yunfei Shao, Jiahui Wu, Qingcheng Wang
article en
4 citations

Abstract

Precise administration of tirofiban must be carefully considered to achieve the best treatment efficacy and maximize safety for patients. Herein, a paper spray ionization (PSI) linear ion trap (LIT) portable mass spectrometer (pMS) based point-of-care testing (POCT) technique was developed for on-site sampling, clinical testing, and immediate analysis of tirofiban blood drug concentrations. The results showed that tirofiban formed a significant and stable parent ion peak at <i>m</i>/<i>z</i> 441.3 by MS1 full scan in positive ion mode, which fragmented into product ions at <i>m</i>/<i>z</i> 395.4, <i>m</i>/<i>z</i> 321.3, <i>m</i>/<i>z</i> 276.3, and <i>m</i>/<i>z</i> 260.3 through collision-induced dissociation (CID) in MS/MS. To improve ion response, the parameters were optimized to the voltages of ionization 3600 V, ion isolation 1 (ISO1) 8 V, ion isolation 2 (ISO2) 2 V, and collision-induced dissociation (CID) 3 V. This method was validated, and the limit of detection (LOD) and limit of quantification (LOQ) were found to be 10.1 and 33.7 μg·L<sup>-1</sup>, respectively. For precision, it had the relative standard deviation (RSD) of interday precision of 4.8 to 6.7% and the RSD of intraday precision of 7.8 to 8.3%. The recovery of the method ranged from 87.5 to 93.4%. Although matrix effects in blood samples had some inhibitory effects on the target signal formation, the method compensated for part of the matrix effects by establishing a matrix-matched calibration curve, which exhibited good linearity with a <i>R</i>2 of 0.9987. Finally, the method was applied to the detection of tirofiban in clinically collected blood samples. Out of 12 samples, ten had tirofiban concentrations between 35.4 and 72.1 μg·L<sup>-1</sup> while the remaining two were below the LOQ. The method needs further optimization and validation in the future to improve its sensitivity and stability.

ACS OmegaVol. 9(51)
Ningbo University (CN), Ministry of Public Security of the People's Republic of China (CN), Nantong University (CN), Wenzhou Medical University (CN), The Affiliated Taizhou People's Hospital of Nanjing Medical University (CN), Institute of Forensic Science (CN), Affiliated Hangzhou First People's Hospital, Westlake University, School of Medicine (CN), Zhejiang Gongshang University (CN), Zhejiang University (CN)
National Natural Science Foundation of China, Zhejiang Medical Association
Good health and well-being
Openalex Percentile: Top 14%
Acute Myocardial Infarction Research
1.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.