Validation Against Polysomnography of a Transthoracic Impedance Sensor for Screening of Sleep Apnea in Heart Failure Patients: A Pooled Analysis of AIRLESS and UPGRADE

<b>Background/Introduction:</b> Cardiac implantable electronic devices and their integrated thoracic impedance sensors have been used to detect sleep apnea for over a decade now. Despite their usage in daily clinical practice, there are only limited data on their diagnostic accuracy. <b>Methods:</b> AIRLESS and UPGRADE were prospective investigator-driven trials meant to validate the AP scan<sup>®</sup> (Boston Scientific, Marlborough, MA, USA) in heart failure cohorts. Patients, who either fulfilled the criteria for implantation of an implantable cardioverter-defibrillator (ICD), cardiac resynchronization therapy (CRT), or upgrading to CRT according to most recent guidelines at the time of study conduction, were eligible for enrolment. Sleep apnea and its severity, measured by apnea-hypopnea index (AHI), were assessed by polysomnography. For direct comparison, the apnea sensor-derived AP scan<sup>®</sup> was used from the identical night. <b>Results:</b> Overall, 80 patients were analyzed. Median AHI was 21.6 events/h (7.1-34.7), while median AP scan<sup>®</sup> was 33.0 events/h (26.0-43.0). In the overall cohort, the sensor-derived AP scan<sup>®</sup> correlated significantly with the AHI (r = 0.61, <i>p</i> < 0.001) with a mean difference (MD) of -12.6 (95% confidence interval (CI) -38.2 to 13.0). Furthermore, the AP scan<sup>®</sup> was found to correlate well with the AHI in patients with obstructive sleep apnea r = 0.73, <i>p</i> = 0.011, MD -5.2, 95% CI -22.7 to 12.3), but not central sleep apnea (r = 0.28, <i>p</i> = 0.348, MD -10.4, 95% CI -35.4 to 14.6). <b>Conclusions:</b> In an exclusive heart failure cohort, the AP scan<sup>®</sup> correlated well with the PSG-derived AHI. A similar correlation was found in most subgroups except for patients suffering from central sleep apnea.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2024-12-10
DOI
https://doi.org/10.3390/jcm13247519
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation Against Polysomnography of a Transthoracic Impedance Sensor for Screening of Sleep Apnea in Heart Failure Patients: A Pooled Analysis of AIRLESS and UPGRADE

Wolfgang Dichtl, Raoua Ben Messaoud, Philipp Spitaler, Peggy Jacon et al.
1 links
Journal of Clinical Medicine
Cardiac pacing and defibrillation studies
article

Validation Against Polysomnography of a Transthoracic Impedance Sensor for Screening of Sleep Apnea in Heart Failure Patients: A Pooled Analysis of AIRLESS and UPGRADE

Wolfgang Dichtl, Raoua Ben Messaoud, Philipp Spitaler, Peggy Jacon, Sabrina Neururer, Thomas Senoner, Sandrine Venier, Fabian Barbieri, Jean‐Louis Pépin, Agnė Adukauskaitė, Florian Hintringer, Bernhard Pfeifer, Pascal Defaye, S. Vergara Limón
article en

Abstract

<b>Background/Introduction:</b> Cardiac implantable electronic devices and their integrated thoracic impedance sensors have been used to detect sleep apnea for over a decade now. Despite their usage in daily clinical practice, there are only limited data on their diagnostic accuracy. <b>Methods:</b> AIRLESS and UPGRADE were prospective investigator-driven trials meant to validate the AP scan<sup>®</sup> (Boston Scientific, Marlborough, MA, USA) in heart failure cohorts. Patients, who either fulfilled the criteria for implantation of an implantable cardioverter-defibrillator (ICD), cardiac resynchronization therapy (CRT), or upgrading to CRT according to most recent guidelines at the time of study conduction, were eligible for enrolment. Sleep apnea and its severity, measured by apnea-hypopnea index (AHI), were assessed by polysomnography. For direct comparison, the apnea sensor-derived AP scan<sup>®</sup> was used from the identical night. <b>Results:</b> Overall, 80 patients were analyzed. Median AHI was 21.6 events/h (7.1-34.7), while median AP scan<sup>®</sup> was 33.0 events/h (26.0-43.0). In the overall cohort, the sensor-derived AP scan<sup>®</sup> correlated significantly with the AHI (r = 0.61, <i>p</i> < 0.001) with a mean difference (MD) of -12.6 (95% confidence interval (CI) -38.2 to 13.0). Furthermore, the AP scan<sup>®</sup> was found to correlate well with the AHI in patients with obstructive sleep apnea r = 0.73, <i>p</i> = 0.011, MD -5.2, 95% CI -22.7 to 12.3), but not central sleep apnea (r = 0.28, <i>p</i> = 0.348, MD -10.4, 95% CI -35.4 to 14.6). <b>Conclusions:</b> In an exclusive heart failure cohort, the AP scan<sup>®</sup> correlated well with the PSG-derived AHI. A similar correlation was found in most subgroups except for patients suffering from central sleep apnea.

Journal of Clinical MedicineVol. 13(24)
Innsbruck Medical University (AT), Inserm (FR), Universität Innsbruck (AT), Centre Hospitalier Universitaire de Grenoble (FR), Deutsches Herzzentrum der Charité (DE), UMIT - Private Universität für Gesundheitswissenschaften, Medizinische Informatik und Technik (AT), Tirol Kliniken (AT), Helmholtz-Zentrum Hereon (DE), Charité - Universitätsmedizin Berlin (DE), Université Grenoble Alpes (FR)
Good health and well-being
Openalex Percentile: Top 58%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.