Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment

To confirm the performance improvement of virtual monoenergetic images (VMIs) for iodine contrast tasks in a clinical photon-counting detector CT (PCD CT) using Fourier-based assessment, compared with those in the latest-generation dual-source dual-energy CT (DECT). A water-filled bath with a diameter of 300 mm, which contains rod-shaped phantoms equivalent to diluted iodine (2 and 12 mg/mL), was scanned using PCD CT and DECT at 15, 7.5, and 3 mGy. VMIs were generated without any iterative reconstruction algorithm. Task transfer function (TTF), noise power spectrum (NPS), and slice sensitivity profile were evaluated for VMIs at 70 and 40 keV. The detectability index (d') and the squared system performance function (SPF<sup>2</sup>) calculated by TTF<sup>2</sup>/NPS were compared. At 40 keV, the d' values of PCD CT were higher (percentage increase of 25.7-39.9%) than those of DECT, whereas at 70 keV, the difference was rather small. The SPF<sup>2</sup> values at 40 keV of PCD CT grew notably higher than those of DECT as the spatial frequency increased. The higher SPF<sup>2</sup> values endorsed the lower image noise and the sharper edge of the rod phantom as observed. The d' and SPF<sup>2</sup> in VMIs at 40 keV of PCD CT were notably higher than those of DECT, which endorsed the clinical advantages of PCD CT that had been previously reported in various studies.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2024-12-10
DOI
https://doi.org/10.1007/s13246-024-01499-6
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.18

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment

Satoshi Kobayashi, Katsuhiro Ichikawa, Hiroji Nagata, Hiroki Kawashima et al.
1 citations
Physical and Engineering Sciences in Medicine
Advanced X-ray and CT Imaging
0.18
article

Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment

Satoshi Kobayashi, Katsuhiro Ichikawa, Hiroji Nagata, Hiroki Kawashima, J. Kataoka, M. Arimoto, Ryoichi Yoshida, Takuto Katayama
article en
1 citations

Abstract

To confirm the performance improvement of virtual monoenergetic images (VMIs) for iodine contrast tasks in a clinical photon-counting detector CT (PCD CT) using Fourier-based assessment, compared with those in the latest-generation dual-source dual-energy CT (DECT). A water-filled bath with a diameter of 300 mm, which contains rod-shaped phantoms equivalent to diluted iodine (2 and 12 mg/mL), was scanned using PCD CT and DECT at 15, 7.5, and 3 mGy. VMIs were generated without any iterative reconstruction algorithm. Task transfer function (TTF), noise power spectrum (NPS), and slice sensitivity profile were evaluated for VMIs at 70 and 40 keV. The detectability index (d') and the squared system performance function (SPF<sup>2</sup>) calculated by TTF<sup>2</sup>/NPS were compared. At 40 keV, the d' values of PCD CT were higher (percentage increase of 25.7-39.9%) than those of DECT, whereas at 70 keV, the difference was rather small. The SPF<sup>2</sup> values at 40 keV of PCD CT grew notably higher than those of DECT as the spatial frequency increased. The higher SPF<sup>2</sup> values endorsed the lower image noise and the sharper edge of the rod phantom as observed. The d' and SPF<sup>2</sup> in VMIs at 40 keV of PCD CT were notably higher than those of DECT, which endorsed the clinical advantages of PCD CT that had been previously reported in various studies.

Physical and Engineering Sciences in MedicineVol. 48(1)
Kanazawa University (JP), Waseda University (JP), Tokai University Hospital (JP), Kanazawa Medical University Hospital (JP)
Kanazawa University, Exploratory Research for Advanced Technology
Affordable and clean energy
Openalex Percentile: Top 50%
Advanced X-ray and CT Imaging
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.