Evaluation of the QRS detector using the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database

Purpose. To evaluate the accuracy of the QRS complex detectorof the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database in accordance with the requirements of GOST R IEC60601–2–47–2017, 30324.2.47–2012. Materials and methods. During the tests, the operation of the detector was evaluated using the MIT-BIH Arrhythmia Database, containing 48 records of 30 minutes each. Each selected QRS complex was assigned one of the following types: N (supraventricular complex, normal QRS complex, including bundle branch block), V (ventricular complex), U (complex that experts find difficult to determine) and A (artifact). Rates of sensitivity (Sen), specificity or positive predictive rate (PPR), and overdiagnosis (FPR) were determined. Results. The sensitivity of the QRS complex detector of the Medicom-combi complex for daily monitoring of ECG and blood pressure was 99,98%, and the specificity was 99,87%. The sensitivity of determining ventricular complexes was 99,97%, the specificity of determining ventricular complexes was 99,64%. Overdiagnosis of ventricular complexes does not exceed 0,03%. Conclusion. The Medicom-combi complex for daily monitoring of ECG and blood pressure, designed for long-term ECG monitoring, has sufficient sensitivity and specificity for analyzing Holter ECG recordings, as well as identifying ventricular ectopic complexes.

Authors

Institutions

Publication Details

Journal
Medical alphabet
Published
2024-12-06
DOI
https://doi.org/10.33667/2078-5631-2024-24-43-46
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the QRS detector using the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database

Yu. N. Fedulaev, Ф А Евдокимов, Г. А. Хайретдинова, А. Николаева et al.
Medical alphabet
ECG Monitoring and Analysis
article

Evaluation of the QRS detector using the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database

Yu. N. Fedulaev, Ф А Евдокимов, Г. А. Хайретдинова, А. Николаева, V. A. Gogichaev
article en

Abstract

Purpose. To evaluate the accuracy of the QRS complex detectorof the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database in accordance with the requirements of GOST R IEC60601–2–47–2017, 30324.2.47–2012. Materials and methods. During the tests, the operation of the detector was evaluated using the MIT-BIH Arrhythmia Database, containing 48 records of 30 minutes each. Each selected QRS complex was assigned one of the following types: N (supraventricular complex, normal QRS complex, including bundle branch block), V (ventricular complex), U (complex that experts find difficult to determine) and A (artifact). Rates of sensitivity (Sen), specificity or positive predictive rate (PPR), and overdiagnosis (FPR) were determined. Results. The sensitivity of the QRS complex detector of the Medicom-combi complex for daily monitoring of ECG and blood pressure was 99,98%, and the specificity was 99,87%. The sensitivity of determining ventricular complexes was 99,97%, the specificity of determining ventricular complexes was 99,64%. Overdiagnosis of ventricular complexes does not exceed 0,03%. Conclusion. The Medicom-combi complex for daily monitoring of ECG and blood pressure, designed for long-term ECG monitoring, has sufficient sensitivity and specificity for analyzing Holter ECG recordings, as well as identifying ventricular ectopic complexes.

Medical alphabet(24)
Pirogov Russian National Research Medical University (RU)
Good health and well-being
Openalex Percentile: Top 59%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of the QRS detector using the Medicom-combi complex for daily monitoring of ECG and blood pressure using the MIT-BIH ECG signal test database — Yu. N. Fedulaev, Ф А Евдокимов, et al. · Medical alphabet (2024) | TGRS Research Map | TGRS