On estimating the effective sensitivity of a TOF-PET system for a particular imaging task

Abstract Comparing positron emission tomography (PET) systems which have different features is not straightforward. To address this, we propose to image the same object with all considered PET systems using a fixed scan time, and reconstruct from each scan an image at the same predefined spatial resolution. With such resolution matched reconstructions, the images should be identical except for their noise. Therefore, the PET system that produces the image with the lowest variance has the best performance. An analytical model is described to compute this variance, assuming that the PET system is (approximately) cylindrical, the imaged object is a uniform cylinder centered in the field of view, and the variance is only computed at the center of the reconstructed image. The model takes into account the solid angle covered by the detectors, the detector stopping power, the time-of-flight (TOF) resolution, the scatter fraction and the spatial resolution of the system, the attenuation and diameter of the cylinder and the desired spatial resolution of the reconstructed image. The inverse of this variance can be considered as the effective sensitivity of the system. This effective sensitivity can be calibrated based on the NEMA line source sensitivity. As a performance metric for scanning long objects, the minimum sensitivity achieved over the object is computed, with an optimal number of bed positions and optimal overlap between them. The influence of the spatial and TOF resolution on the effective sensitivity is verified with simulations.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2024-12-01
DOI
https://doi.org/10.1088/1748-0221/19/12/c12002
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On estimating the effective sensitivity of a TOF-PET system for a particular imaging task

S. Abdollahi Pour, S. Noë, Georg Schramm, Johan Nuyts et al.
Journal of Instrumentation
Medical Imaging Techniques and Applications
article

On estimating the effective sensitivity of a TOF-PET system for a particular imaging task

S. Abdollahi Pour, S. Noë, Georg Schramm, Johan Nuyts, Ahmadreza Rezaei
article en

Abstract

Abstract Comparing positron emission tomography (PET) systems which have different features is not straightforward. To address this, we propose to image the same object with all considered PET systems using a fixed scan time, and reconstruct from each scan an image at the same predefined spatial resolution. With such resolution matched reconstructions, the images should be identical except for their noise. Therefore, the PET system that produces the image with the lowest variance has the best performance. An analytical model is described to compute this variance, assuming that the PET system is (approximately) cylindrical, the imaged object is a uniform cylinder centered in the field of view, and the variance is only computed at the center of the reconstructed image. The model takes into account the solid angle covered by the detectors, the detector stopping power, the time-of-flight (TOF) resolution, the scatter fraction and the spatial resolution of the system, the attenuation and diameter of the cylinder and the desired spatial resolution of the reconstructed image. The inverse of this variance can be considered as the effective sensitivity of the system. This effective sensitivity can be calibrated based on the NEMA line source sensitivity. As a performance metric for scanning long objects, the minimum sensitivity achieved over the object is computed, with an optimal number of bed positions and optimal overlap between them. The influence of the spatial and TOF resolution on the effective sensitivity is verified with simulations.

Journal of InstrumentationVol. 19(12)
KU Leuven (BE)
Fonds Wetenschappelijk Onderzoek, Vlaamse regering
Openalex Percentile: Top 61%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.