X-ray–induced acoustic computed tomography: 3D X-ray absorption imaging from a single view

Computed tomography (CT) scanners are essential for modern imaging but require around 600 projections from various angles. We present x-ray-induced acoustic computed tomography (XACT), a method that uses radiation-induced acoustic waves for three-dimensional (3D) x-ray imaging. These spherical acoustic waves travel through tissue at 1.5 × 10<sup>3</sup> meters per second, much slower than x-rays, allowing ultrasound detectors to capture them and generate 3D images without mechanical scanning. We validate this theory by performing 3D numerical reconstructions of a human breast from a single x-ray projection and experimentally determining 3D structures of objects at different depths. Achieving resolutions of 0.4 millimeters in the <i>XZ</i> plane and 3.5 millimeters in the <i>XY</i> plane at a depth of 16 millimeters, XACT demonstrates the ability to produce 3D images from one x-ray projection, reducing radiation exposure and enabling gantry-free imaging. XACT shows great promise for biomedical and nondestructive testing applications, potentially replacing conventional CT.

Authors

Institutions

Publication Details

Journal
Science Advances
Published
2024-12-06
DOI
https://doi.org/10.1126/sciadv.ads1584
Citations
10
Primary Topic
Photoacoustic and Ultrasonic Imaging
Type
article
Field-Weighted Citation Impact
1.84

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

X-ray–induced acoustic computed tomography: 3D X-ray absorption imaging from a single view

Gerald Lee, Prabodh Kumar Pandey, Leshan Sun, Liangzhong Xiang et al.
10 citations
Science Advances
Photoacoustic and Ultrasonic Imaging
1.84
article

X-ray–induced acoustic computed tomography: 3D X-ray absorption imaging from a single view

Gerald Lee, Prabodh Kumar Pandey, Leshan Sun, Liangzhong Xiang, Rick J. P. van Bergen, Siqi Wang, Yifei Xu
article en
10 citations

Abstract

Computed tomography (CT) scanners are essential for modern imaging but require around 600 projections from various angles. We present x-ray-induced acoustic computed tomography (XACT), a method that uses radiation-induced acoustic waves for three-dimensional (3D) x-ray imaging. These spherical acoustic waves travel through tissue at 1.5 × 10<sup>3</sup> meters per second, much slower than x-rays, allowing ultrasound detectors to capture them and generate 3D images without mechanical scanning. We validate this theory by performing 3D numerical reconstructions of a human breast from a single x-ray projection and experimentally determining 3D structures of objects at different depths. Achieving resolutions of 0.4 millimeters in the <i>XZ</i> plane and 3.5 millimeters in the <i>XY</i> plane at a depth of 16 millimeters, XACT demonstrates the ability to produce 3D images from one x-ray projection, reducing radiation exposure and enabling gantry-free imaging. XACT shows great promise for biomedical and nondestructive testing applications, potentially replacing conventional CT.

Science AdvancesVol. 10(49)
University of California, Irvine (US), Beckman Laser Institute and Medical Clinic (US)
National Institutes of Health, Chao Family Comprehensive Cancer Center
Openalex Percentile: Top 15%
Photoacoustic and Ultrasonic Imaging
1.84
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.