A 16-Output 10-V Compliant Stimulator ASIC With Sub-10-nA Mismatch and Simultaneous ETI Sensing for Selective Neural Stimulation

This article presents a 16-output high-voltage (HV) compliant stimulator ASIC for selective neural stimulation. The ASIC supports temporal interference stimulation (TIS) to achieve high-spatial selectivity without requiring nerve-penetrating electrodes. A novel on-the-fly active charge balancing (CB) is proposed since existing CB solutions cannot be directly applied to TIS. Simultaneous electrode-tissue impedance (ETI) sensing is realized by reusing the CB hardware. The stimulator ASIC, fabricated in 130-nm bipolar-complementary metal-oxide–semiconductor-double-diffused metal-oxide–semiconductor (BCD) technology, occupies 0.29 mm2 per output and achieves 10-V compliance while supporting up to 10-mA stimulation current and maintaining TIS steering flexibility. The proposed active CB approach compensates for electrode voltage drift during TIS based on a negative feedback loop, achieving a sub-10-nA mismatch current over a wide range of ETIs. The ETI sensing reuses the stimulation current and CB hardware for simultaneous measurements during stimulation, achieving a sensing inaccuracy of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\pm 2~\Omega $ </tex-math></inline-formula>. Extensive saline experiments confirm the ability of the ASIC to achieve superior spatial selectivity for stimulation while maintaining proper active CB and simultaneous ETI sensing.

Authors

Institutions

Publication Details

Journal
IEEE Journal of Solid-State Circuits
Published
2024-12-04
DOI
https://doi.org/10.1109/jssc.2024.3503914
Citations
4
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
1.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A 16-Output 10-V Compliant Stimulator ASIC With Sub-10-nA Mismatch and Simultaneous ETI Sensing for Selective Neural Stimulation

Haoming Xin, Mario Konijnenburg, Carolina Mora López, Nick Van Helleputte et al.
4 citations
IEEE Journal of Solid-State Circuits
Neuroscience and Neural Engineering
1.07
article

A 16-Output 10-V Compliant Stimulator ASIC With Sub-10-nA Mismatch and Simultaneous ETI Sensing for Selective Neural Stimulation

Haoming Xin, Mario Konijnenburg, Carolina Mora López, Nick Van Helleputte, Shrishail Patki, Nicolò Rossetti, Konstantinos Petkos, Peter Vis, Geert Langereis, Roland van Wegberg, Vojkan Mihajlović, Meiyi Zhou, Mark Fichman
article en
4 citations

Abstract

This article presents a 16-output high-voltage (HV) compliant stimulator ASIC for selective neural stimulation. The ASIC supports temporal interference stimulation (TIS) to achieve high-spatial selectivity without requiring nerve-penetrating electrodes. A novel on-the-fly active charge balancing (CB) is proposed since existing CB solutions cannot be directly applied to TIS. Simultaneous electrode-tissue impedance (ETI) sensing is realized by reusing the CB hardware. The stimulator ASIC, fabricated in 130-nm bipolar-complementary metal-oxide–semiconductor-double-diffused metal-oxide–semiconductor (BCD) technology, occupies 0.29 mm2 per output and achieves 10-V compliance while supporting up to 10-mA stimulation current and maintaining TIS steering flexibility. The proposed active CB approach compensates for electrode voltage drift during TIS based on a negative feedback loop, achieving a sub-10-nA mismatch current over a wide range of ETIs. The ETI sensing reuses the stimulation current and CB hardware for simultaneous measurements during stimulation, achieving a sensing inaccuracy of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\pm 2~\Omega $ </tex-math></inline-formula>. Extensive saline experiments confirm the ability of the ASIC to achieve superior spatial selectivity for stimulation while maintaining proper active CB and simultaneous ETI sensing.

IEEE Journal of Solid-State CircuitsVol. 60(3)
Imec the Netherlands (NL), IMEC (BE)
Openalex Percentile: Top 23%
Neuroscience and Neural Engineering
1.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.