Compact high-bandwidth single-beam optically-pumped magnetometer for biomagnetic measurement
Optically-pumped magnetometer (OPM) has been of increasing interest for biomagnetic measurements due to its low cost and portability compared with superconducting quantum interference devices (SQUID). Miniaturized spin-exchange-relaxation-free (SERF) OPMs typically have limited bandwidth (less than a few hundred Hertz), making it difficult to measure high-frequency biomagnetic signals such as the magnetocardiography (MCG) signal of the mouse. Existing experiments mainly use SQUID systems to measure the signal. In this paper, we introduce a prototype miniaturized single-beam SERF magnetometer with a bandwidth of ∼ 1 kHz. Instead of operating the OPM in a closed-loop mode to improve the bandwidth of the OPM, which usually has a poorer performance in high-frequency range, we use the power-broadening effects to shorten the spin relaxation time and thus a faster response to the magnetic fields to be measured. Combined with light power stabilizations to improve both the sensitivity and stability, our magnetometer has a low noise floor of 30 fT / Hz<sup>1/2</sup>, which has been successfully adopted to measure the MCG signal of the mouse.
Authors
- Xiaoyong Wei (ORCID: https://orcid.org/0000-0001-9677-4227)
- Xiang Peng (ORCID: https://orcid.org/0000-0003-0815-5122)
- Teng Wu (ORCID: https://orcid.org/0000-0002-7707-7474)
- Hong Guo (ORCID: https://orcid.org/0000-0003-0644-6698)
- T.‐C. Wu
Institutions
- Beijing Institute of Technology (CN)
- Peking University (CN)
- California Northstate University (US)
- Ministry of Industry and Information Technology (CN)
Publication Details
- Journal
- Biomedical Optics Express
- Published
- 2024-12-04
- DOI
- https://doi.org/10.1364/boe.545624
- Citations
- 7
- Primary Topic
- Atomic and Subatomic Physics Research
- Type
- article
- Field-Weighted Citation Impact
- 2.36
Funders
- National Natural Science Foundation of China
- Peking University