Analisis Hubungan Nilai Noise Citra dan Computed Tomography Dose Index (CTDI) Pada Variasi Pitch Protokol CT Abdomen Rutin di RSUD Koja Jakarta Utara

Computed Tomography (CT) has become one of the primary diagnostic methods in modern medicine, allowing detailed and accurate visualization of the internal structures of the human body. In its use, the pitch value can affect the radiation dose received by the patient and the quality of the resulting image. This study aims to analyze the effect of pitch variations on the Computed Tomography Dose Index (CTDI) and image noise in Routine Abdominal CT to achieve the best image quality with minimal patient dose. This research method uses a quantitative design using a 32 cm diameter PMMA water phantom as a model. Eight pitch variations (0.40, 0.55, 0.60, 0.80, 0.90, 1.00, 1.35, and 1.50) were evaluated using a Siemens Somatom Perspective 128 Slice CT-Scan. Data were collected through experiments on the water phantom and analyzed using descriptive analysis, normality tests, and correlation tests. The results showed no significant relationship between CTDI values and image noise with pitch variations. Increasing the pitch value contributed to increasing CTDI but decreased image noise. These results indicate that a pitch of 0.60 has an optimal image noise value and minimal CTDI in the Routine CT Abdomen protocol at RSUD Koja, North Jakarta. This study provides a deeper understanding of the importance of selecting the correct pitch value in minimizing patient radiation dose while maintaining good image quality. The practical implication is using a pitch of 0.60 as the optimal value for the Routine CT Abdomen protocol, which can reduce the risk of excessive radiation exposure to patients. Keywords: CT-Scan, pitch, water phantom, CT Abdomen, noise, computed tomography dose index (CTDI)

Authors

Publication Details

Journal
Jurnal Pengawasan Tenaga Nuklir
Published
2024-12-04
DOI
https://doi.org/10.53862/jupeten.v4i2.012
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analisis Hubungan Nilai Noise Citra dan Computed Tomography Dose Index (CTDI) Pada Variasi Pitch Protokol CT Abdomen Rutin di RSUD Koja Jakarta Utara

Wahyu Hidayat, Eka Putra Syarif Hidayat, Nursama Heru, Muhammad Rizqi et al.
Jurnal Pengawasan Tenaga Nuklir
Advanced X-ray and CT Imaging
article

Analisis Hubungan Nilai Noise Citra dan Computed Tomography Dose Index (CTDI) Pada Variasi Pitch Protokol CT Abdomen Rutin di RSUD Koja Jakarta Utara

Wahyu Hidayat, Eka Putra Syarif Hidayat, Nursama Heru, Muhammad Rizqi, Nurbaiti Nurbaiti, Kamila Kamila, Khaeroel Ansory
article en

Abstract

Computed Tomography (CT) has become one of the primary diagnostic methods in modern medicine, allowing detailed and accurate visualization of the internal structures of the human body. In its use, the pitch value can affect the radiation dose received by the patient and the quality of the resulting image. This study aims to analyze the effect of pitch variations on the Computed Tomography Dose Index (CTDI) and image noise in Routine Abdominal CT to achieve the best image quality with minimal patient dose. This research method uses a quantitative design using a 32 cm diameter PMMA water phantom as a model. Eight pitch variations (0.40, 0.55, 0.60, 0.80, 0.90, 1.00, 1.35, and 1.50) were evaluated using a Siemens Somatom Perspective 128 Slice CT-Scan. Data were collected through experiments on the water phantom and analyzed using descriptive analysis, normality tests, and correlation tests. The results showed no significant relationship between CTDI values and image noise with pitch variations. Increasing the pitch value contributed to increasing CTDI but decreased image noise. These results indicate that a pitch of 0.60 has an optimal image noise value and minimal CTDI in the Routine CT Abdomen protocol at RSUD Koja, North Jakarta. This study provides a deeper understanding of the importance of selecting the correct pitch value in minimizing patient radiation dose while maintaining good image quality. The practical implication is using a pitch of 0.60 as the optimal value for the Routine CT Abdomen protocol, which can reduce the risk of excessive radiation exposure to patients. Keywords: CT-Scan, pitch, water phantom, CT Abdomen, noise, computed tomography dose index (CTDI)

Jurnal Pengawasan Tenaga NuklirVol. 4(2)
Openalex Percentile: Top 79%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Analisis Hubungan Nilai Noise Citra dan Computed Tomography Dose Index (CTDI) Pada Variasi Pitch Protokol CT Abdomen Rutin di RSUD Koja Jakarta Utara — Wahyu Hidayat, Eka Putra Syarif Hidayat, et al. · Jurnal Pengawasan Tenaga Nuklir (2024) | TGRS Research Map | TGRS