Accuracy of volume measurements by a clinical spirometer in multiple veterinary hospitals
These results support caution when interpreting clinical measurements of lung volumes and mechanics in anaesthetised patients when using these sensors. This is particularly important in smaller patients where lung volumes are below 300 mL. Trends should be reliable.
Authors
- Olivier Levionnois (ORCID: https://orcid.org/0000-0003-0253-4682)
- Adam Auckburally (ORCID: https://orcid.org/0000-0002-3702-2145)
- Mathieu Raillard (ORCID: https://orcid.org/0000-0003-4057-9312)
- J. Soares (ORCID: https://orcid.org/0000-0001-9044-3208)
- Tristan Merlin (ORCID: https://orcid.org/0000-0002-0996-4629)
- Barbara Steblaj (ORCID: https://orcid.org/0000-0003-3812-6103)
- Frances Downing
- Alejandra García de Carellán Mateo (ORCID: https://orcid.org/0000-0002-5725-9991)
- Paul MacFarlane (ORCID: https://orcid.org/0000-0002-3167-4855)
- Karla Borland (ORCID: https://orcid.org/0000-0002-5851-6101)
- Susana Canfrán (ORCID: https://orcid.org/0000-0003-0104-2010)
- Karine Portier (ORCID: https://orcid.org/0000-0002-0200-3462)
- Martina Mosing (ORCID: https://orcid.org/0000-0002-6190-5642)
- Aurora Zoff
- William McFadzean
- Anthea L. Raisis
- Josephine Robertson
Institutions
- Université Claude Bernard Lyon 1 (FR)
- University of Bern (CH)
- Universidad Complutense de Madrid (ES)
- University of Veterinary Medicine Vienna (AT)
- Inserm (FR)
- Murdoch University (AU)
- University of Zurich (CH)
- Valencia Catholic University Saint Vincent Martyr (ES)
- University of Bristol (GB)
- Centre de Recherche en Neurosciences de Lyon (FR)
- VetAgro Sup (FR)
- North Downs Hospital (GB)
- University of Glasgow (GB)
- University of California, Davis (US)
Publication Details
- Journal
- Frontiers in Veterinary Science
- Published
- 2024-12-04
- DOI
- https://doi.org/10.3389/fvets.2024.1475401
- Primary Topic
- Hemodynamic Monitoring and Therapy
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Australian Government