Alignment requirement of retro-reflective target in self-mixing interferometry
It is generally assumed that using a retro-reflective surface as the target of an interferometric measurement, an accurate alignment to the illumination beam is not really necessary. On the contrary, we show that even a few degrees of angular misalignment can corrupt the interferometric signal with the onset of a large artifact, which is also misleading because it is similar to the waveform of the high-level feedback regime. We first provide the experimental evidence of the artifact in a self-mixing interferometer, showing that the artifact is canceled after an alignment to <0.2 deg error, and then provide a theoretical evaluation, in good agreement with the observations.
Authors
- Adrien Albert (ORCID: https://orcid.org/0000-0002-0893-0881)
- Silvano Donati (ORCID: https://orcid.org/0000-0002-2977-0194)
- San‐Liang Lee (ORCID: https://orcid.org/0000-0003-2436-0273)
Institutions
- National Taiwan University of Science and Technology (TW)
- University of Pavia (IT)
Publication Details
- Journal
- Optical Engineering
- Published
- 2024-12-04
- DOI
- https://doi.org/10.1117/1.oe.63.12.124102
- Primary Topic
- Semiconductor Lasers and Optical Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.00