Agreement between Two Biometers for Central Corneal Thickness Measurement in High Myopia Patients

Background In refractive surgery, the measurement of Central Corneal Thickness (CCT) is a critical factor in the preparation and screening of surgery to correct myopia. CCT measurements can be measured by optical biometers, including the Scheimpflug camera and Swept-source Optical Coherence Tomography (SS-OCT). Aim This study aimed to compare the Central Corneal Thickness (CCT) measurements made with Swept-source Optical Coherence Tomography (SS-OCT)-based biometers and the Scheimpflug camera in patients with high myopia. Methods An analytical cross-sectional study was performed to evaluate the agreement between Central Corneal Thickness (CCT) values obtained using Swept-source Optical Coherence Tomography (SS-OCT) and the Scheimpflug camera. We included a total of 28 individuals with high myopia, including 44 eyes, for our study. These patients underwent Central Corneal Thickness (CCT) tests using both Swept-source Optical Coherence Tomography (SS-OCT) and the Scheimpflug camera. The Bland-Altman test was used to assess the agreement between the two devices. Results Forty-four eyes with axial lengths ranging from 26.05 to 34.02 mm based on SS-OCT examinations were included in this study. The mean CCT was 527.59 ± 35.710 ( p <0.05) μm and 525.64 ± 35.421 μm ( p <0.05), based on SS-OCT and the Scheimpflug camera, respectively. The Bland-Altman plot revealed good agreement between both devices, with a mean difference of 2.3409 μm; however, a 95% limit of agreement demonstrated a wide range, ranging from -8.48509 to 12.39400 μm at a 95% confidence interval. Conclusion CCT measurements using SS-OCT and the Scheimpflug camera differed significantly among patients with high myopia. Our findings have suggested that CCT measurements made with either device should be interpreted with caution and not be used interchangeably.

Authors

Institutions

Publication Details

Journal
The Open Ophthalmology Journal
Published
2024-12-03
DOI
https://doi.org/10.2174/0118743641344548241105115415
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement between Two Biometers for Central Corneal Thickness Measurement in High Myopia Patients

Budiman Budiman, Emmy Dwi Sugiarti, Ludwig Melino Tjokrovonco, Andrew M. H. Knoch
The Open Ophthalmology Journal
Corneal surgery and disorders
article

Agreement between Two Biometers for Central Corneal Thickness Measurement in High Myopia Patients

Budiman Budiman, Emmy Dwi Sugiarti, Ludwig Melino Tjokrovonco, Andrew M. H. Knoch
article en

Abstract

Background In refractive surgery, the measurement of Central Corneal Thickness (CCT) is a critical factor in the preparation and screening of surgery to correct myopia. CCT measurements can be measured by optical biometers, including the Scheimpflug camera and Swept-source Optical Coherence Tomography (SS-OCT). Aim This study aimed to compare the Central Corneal Thickness (CCT) measurements made with Swept-source Optical Coherence Tomography (SS-OCT)-based biometers and the Scheimpflug camera in patients with high myopia. Methods An analytical cross-sectional study was performed to evaluate the agreement between Central Corneal Thickness (CCT) values obtained using Swept-source Optical Coherence Tomography (SS-OCT) and the Scheimpflug camera. We included a total of 28 individuals with high myopia, including 44 eyes, for our study. These patients underwent Central Corneal Thickness (CCT) tests using both Swept-source Optical Coherence Tomography (SS-OCT) and the Scheimpflug camera. The Bland-Altman test was used to assess the agreement between the two devices. Results Forty-four eyes with axial lengths ranging from 26.05 to 34.02 mm based on SS-OCT examinations were included in this study. The mean CCT was 527.59 ± 35.710 ( p <0.05) μm and 525.64 ± 35.421 μm ( p <0.05), based on SS-OCT and the Scheimpflug camera, respectively. The Bland-Altman plot revealed good agreement between both devices, with a mean difference of 2.3409 μm; however, a 95% limit of agreement demonstrated a wide range, ranging from -8.48509 to 12.39400 μm at a 95% confidence interval. Conclusion CCT measurements using SS-OCT and the Scheimpflug camera differed significantly among patients with high myopia. Our findings have suggested that CCT measurements made with either device should be interpreted with caution and not be used interchangeably.

The Open Ophthalmology JournalVol. 18(1)
Dr. Hasan Sadikin General Hospital (ID), Padjadjaran University (ID)
Openalex Percentile: Top 62%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.