An Open-Frame Loading Stage for High-Resolution X-Ray CT

The utilisation of high-resolution in situ computed tomography (CT) in the (sub-)μm range is typically only viable in synchrotron facilities, as the deployment of a conventional loading stage in laboratory CTs with a cone beam source does not facilitate a corresponding geometric magnification. This publication presents a CT system with a novel in situ concept that allows spatial resolutions down to 0.5 μm, enabling the analysis of weakly absorbing materials capable of applying loads of up to 5 kN in both the compression and tension directions to the sample during the measurement. The necessity for a highly precise mechanical design to ensure successful measurements at magnifications approaching the theoretical limit makes the system’s development particularly demanding. The components employed are presented, along with the requisite considerations and methodologies. It can be demonstrated that the intended specifications with regard to precision and quality are met. The experimental results of a fibre-reinforced polymer demonstrate the system’s ability to detect matrix damage features below a single fibre diameter, thereby highlighting its potential for applications in materials science where traditional laboratory CTs are insufficient and synchrotron access is limited.

Authors

Institutions

Publication Details

Journal
Instruments
Published
2024-12-03
DOI
https://doi.org/10.3390/instruments8040052
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Open-Frame Loading Stage for High-Resolution X-Ray CT

Mário F. M. Campos, Stefan Hiermaier, Georg C. Ganzenmueller, M. Fiederle et al.
1 citations
Instruments
Medical Imaging Techniques and Applications
0.40
article

An Open-Frame Loading Stage for High-Resolution X-Ray CT

Mário F. M. Campos, Stefan Hiermaier, Georg C. Ganzenmueller, M. Fiederle, Frank Fischer, David Plappert, S. Procz, M. K. Schütz
article en
1 citations

Abstract

The utilisation of high-resolution in situ computed tomography (CT) in the (sub-)μm range is typically only viable in synchrotron facilities, as the deployment of a conventional loading stage in laboratory CTs with a cone beam source does not facilitate a corresponding geometric magnification. This publication presents a CT system with a novel in situ concept that allows spatial resolutions down to 0.5 μm, enabling the analysis of weakly absorbing materials capable of applying loads of up to 5 kN in both the compression and tension directions to the sample during the measurement. The necessity for a highly precise mechanical design to ensure successful measurements at magnifications approaching the theoretical limit makes the system’s development particularly demanding. The components employed are presented, along with the requisite considerations and methodologies. It can be demonstrated that the intended specifications with regard to precision and quality are met. The experimental results of a fibre-reinforced polymer demonstrate the system’s ability to detect matrix damage features below a single fibre diameter, thereby highlighting its potential for applications in materials science where traditional laboratory CTs are insufficient and synchrotron access is limited.

InstrumentsVol. 8(4)
University of Freiburg (DE), Fraunhofer Institute for High-Speed Dynamics, Ernst-Mach-Institut (DE)
Openalex Percentile: Top 32%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

An Open-Frame Loading Stage for High-Resolution X-Ray CT — Mário F. M. Campos, Stefan Hiermaier, et al. · Instruments (2024) | TGRS Research Map | TGRS