The Mechanism and Management of Pneumopericardium Caused by Right Ventricular Lead Perforation

An 83-year-old man underwent dual-chamber pacemaker placement for complete atrioventricular block at another hospital. The active-fixation ventricular lead was positioned on the free wall of the anterior right ventricle. Ventricular pacing failure occurred on the day after pacemaker implantation, and fluoroscopy revealed right ventricular (RV) lead perforation. The patient was transferred to our hospital, and chest computed tomography revealed a severe pneumothorax and moderate pneumopericardium. These symptoms were relieved after chest tube drainage, and the patient's hemodynamics stabilized. The RV lead was percutaneously removed using simple traction under fluoroscopic guidance with cardiac surgical backup and was uneventfully refixed to the RV septum. Although there have been several reports of pneumopericardium caused by atrial lead perforation, there are very few cases related to RV lead. Pneumopericardium complicated by pneumothorax due to RV lead perforation can be relieved using chest tube drainage without the need for pericardiocentesis.

Authors

Institutions

Publication Details

Journal
Cardiology Research
Published
2024-12-01
DOI
https://doi.org/10.14740/cr1738
Primary Topic
Pneumothorax, Barotrauma, Emphysema
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Mechanism and Management of Pneumopericardium Caused by Right Ventricular Lead Perforation

Yuichiro Ueno, Noriyuki Mohri, Akihito Ideishi, Tomo Komaki et al.
Cardiology Research
Pneumothorax, Barotrauma, Emphysema
article

The Mechanism and Management of Pneumopericardium Caused by Right Ventricular Lead Perforation

Yuichiro Ueno, Noriyuki Mohri, Akihito Ideishi, Tomo Komaki, Masahiro Ogawa, Shin-ichiro Miura, Kohei Tashiro
article en

Abstract

An 83-year-old man underwent dual-chamber pacemaker placement for complete atrioventricular block at another hospital. The active-fixation ventricular lead was positioned on the free wall of the anterior right ventricle. Ventricular pacing failure occurred on the day after pacemaker implantation, and fluoroscopy revealed right ventricular (RV) lead perforation. The patient was transferred to our hospital, and chest computed tomography revealed a severe pneumothorax and moderate pneumopericardium. These symptoms were relieved after chest tube drainage, and the patient's hemodynamics stabilized. The RV lead was percutaneously removed using simple traction under fluoroscopic guidance with cardiac surgical backup and was uneventfully refixed to the RV septum. Although there have been several reports of pneumopericardium caused by atrial lead perforation, there are very few cases related to RV lead. Pneumopericardium complicated by pneumothorax due to RV lead perforation can be relieved using chest tube drainage without the need for pericardiocentesis.

Cardiology ResearchVol. 15(6)
Fukuoka University (JP), Fukuoka University Hospital (JP)
Openalex Percentile: Top 59%
Pneumothorax, Barotrauma, Emphysema
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The Mechanism and Management of Pneumopericardium Caused by Right Ventricular Lead Perforation — Yuichiro Ueno, Noriyuki Mohri, et al. · Cardiology Research (2024) | TGRS Research Map | TGRS