Effect of the oblique incidence of radiation beams on emerging radiation behind lead and concrete shields: a multilayer method for dose transmission calculations

<i>Objective</i>. For calculating shielding in x-ray rooms, it is often assumed that the beams impinge perpendicularly on the protective barriers. This is not always true, but this premise simplifies the calculations and enhances protection by being a conservative calculation. In this work, a method for calculating radiation transmission through planar shielding that considers the obliquity of the incident beam is presented.<i>Approach</i>. The output of the method produces energy spectra according to the direction of radiation impinging on the shielding. Four angles of incidence on the barrier are considered, along with monoenergetic pencil beams with energies ranging from 10 to 150 keV and two materials: lead and concrete. The direction of emerging photons is discretized into 49 different direction vectors. Monte Carlo calculations are performed for thicknesses of 0.1, 0.5, and 1.0 mm of lead, and 1, 5, 10, and 15 cm of concrete. Additionally, a multilayer iterative method is implemented for calculating attenuation of other thicknesses.<i>Main results</i>. The distribution of radiant energy according to the coordinates of its directional vector illustrates the effect of the obliquity of the incidence and the significance of the shielding material employed. In the case of concrete, the dispersion of radiation away from the original direction of incidence is much more pronounced than in the case of lead at energies below its K-edge. The multilayer iterative method provides highly accurate values of transmitted radiant energy in both monoenergetic and polyenergetic beams, for both lead and concrete, across the various studied incidence directions.<i>Significance</i>. Considering the direction of the photons reaching a shield and the direction of the photons passing through it allows multilayer composite shielding calculations to closely approximate the calculation made for the composite shielding.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-12-03
DOI
https://doi.org/10.1088/1361-6560/ad9a4c
Citations
1
Primary Topic
Graphite, nuclear technology, radiation studies
Type
article
Field-Weighted Citation Impact
0.12
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of the oblique incidence of radiation beams on emerging radiation behind lead and concrete shields: a multilayer method for dose transmission calculations

Antonio González‐López
1 citations
Physics in Medicine and Biology
Graphite, nuclear technology, radiation studies
0.12
article

Effect of the oblique incidence of radiation beams on emerging radiation behind lead and concrete shields: a multilayer method for dose transmission calculations

Antonio González‐López
article en
1 citations

Abstract

<i>Objective</i>. For calculating shielding in x-ray rooms, it is often assumed that the beams impinge perpendicularly on the protective barriers. This is not always true, but this premise simplifies the calculations and enhances protection by being a conservative calculation. In this work, a method for calculating radiation transmission through planar shielding that considers the obliquity of the incident beam is presented.<i>Approach</i>. The output of the method produces energy spectra according to the direction of radiation impinging on the shielding. Four angles of incidence on the barrier are considered, along with monoenergetic pencil beams with energies ranging from 10 to 150 keV and two materials: lead and concrete. The direction of emerging photons is discretized into 49 different direction vectors. Monte Carlo calculations are performed for thicknesses of 0.1, 0.5, and 1.0 mm of lead, and 1, 5, 10, and 15 cm of concrete. Additionally, a multilayer iterative method is implemented for calculating attenuation of other thicknesses.<i>Main results</i>. The distribution of radiant energy according to the coordinates of its directional vector illustrates the effect of the obliquity of the incidence and the significance of the shielding material employed. In the case of concrete, the dispersion of radiation away from the original direction of incidence is much more pronounced than in the case of lead at energies below its K-edge. The multilayer iterative method provides highly accurate values of transmitted radiant energy in both monoenergetic and polyenergetic beams, for both lead and concrete, across the various studied incidence directions.<i>Significance</i>. Considering the direction of the photons reaching a shield and the direction of the photons passing through it allows multilayer composite shielding calculations to closely approximate the calculation made for the composite shielding.

Physics in Medicine and BiologyVol. 69(24)
Hospital Universitario Virgen de la Arrixaca (ES)
Affordable and clean energy
Openalex Percentile: Top 57%
Graphite, nuclear technology, radiation studies
0.12
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Effect of the oblique incidence of radiation beams on emerging radiation behind lead and concrete shields: a multilayer method for dose transmission calculations — Antonio González‐López · Physics in Medicine and Biology (2024) | TGRS Research Map | TGRS