Quantitative neuromuscular monitoring with train-of-four ratio using sonomechanomyography (SMMG): Toward more sensitive evaluation of neuromuscular blockade

In the post-anesthesia care unit, there is a high occurrence of residual neuromuscular blockade, which puts patients at risk of negative consequences such as hypoxia. Assessment based on the train-of-four ratio (TOFR) has been used to avoid residual neuromuscular blockade when the TOFR is greater than 0.9, measured at the adductor pollicis muscle (APM). The most commonly used quantitative neuromuscular monitoring (QNM) modalities include acceleromyography (AMG) and electromyography (EMG). However, the poor user-friendliness of current QNM methods hinders their widespread adoption. To overcome this, we developed a new monitoring method using ultra-fast ultrasound imaging to generate a two-dimensional map of muscle transient motion, i.e., sonomechanomyography (SMMG). SMMG of the APM and AMG of the thumb were used to get the TOFR of 20 normal adults. The results showed no significant difference between the left and right hands for both AMG and SMMG TOFR, with p-values larger than 0.05. In addition, the mean accuracy of SMMG TOFR (0.6% relative error) was higher than AMG (1.4% relative error). Moreover, the Bland-Altman plot showed that all the difference values were within the limits of agreement and the mean bias was 0.02, indicating that the two methods had a very good agreement. In particular, using SMMG did not require additional calibration before testing. Overall, the results demonstrated that the method has the potential as a new QNM approach for further clinical studies to benefit patients in need. To demonstrate its clinical potential, further studies are required to evaluate this method in patients during and post-anesthesia.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2024-12-01
DOI
https://doi.org/10.1063/5.0243459
Citations
1
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative neuromuscular monitoring with train-of-four ratio using sonomechanomyography (SMMG): Toward more sensitive evaluation of neuromuscular blockade

Pancheng Zhu, Samford Wong, Yong‐Ping Zheng, Zhen Song
1 citations
Review of Scientific Instruments
Hemodynamic Monitoring and Therapy
0.57
article

Quantitative neuromuscular monitoring with train-of-four ratio using sonomechanomyography (SMMG): Toward more sensitive evaluation of neuromuscular blockade

Pancheng Zhu, Samford Wong, Yong‐Ping Zheng, Zhen Song
article en
1 citations

Abstract

In the post-anesthesia care unit, there is a high occurrence of residual neuromuscular blockade, which puts patients at risk of negative consequences such as hypoxia. Assessment based on the train-of-four ratio (TOFR) has been used to avoid residual neuromuscular blockade when the TOFR is greater than 0.9, measured at the adductor pollicis muscle (APM). The most commonly used quantitative neuromuscular monitoring (QNM) modalities include acceleromyography (AMG) and electromyography (EMG). However, the poor user-friendliness of current QNM methods hinders their widespread adoption. To overcome this, we developed a new monitoring method using ultra-fast ultrasound imaging to generate a two-dimensional map of muscle transient motion, i.e., sonomechanomyography (SMMG). SMMG of the APM and AMG of the thumb were used to get the TOFR of 20 normal adults. The results showed no significant difference between the left and right hands for both AMG and SMMG TOFR, with p-values larger than 0.05. In addition, the mean accuracy of SMMG TOFR (0.6% relative error) was higher than AMG (1.4% relative error). Moreover, the Bland-Altman plot showed that all the difference values were within the limits of agreement and the mean bias was 0.02, indicating that the two methods had a very good agreement. In particular, using SMMG did not require additional calibration before testing. Overall, the results demonstrated that the method has the potential as a new QNM approach for further clinical studies to benefit patients in need. To demonstrate its clinical potential, further studies are required to evaluate this method in patients during and post-anesthesia.

Review of Scientific InstrumentsVol. 95(12)
Hong Kong Polytechnic University (HK), University of Hong Kong (HK)
No poverty
Openalex Percentile: Top 27%
Hemodynamic Monitoring and Therapy
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.