Clinical evaluation of performance, stability, and longevity of an accelerator system designed for boron neutron capture therapy utilising a beryllium target
Boron neutron capture therapy for recurrent head and neck cancer has been approved as an insurance covered treatment in Japan since June 2020. The Kansai BNCT Medical Center has installed the NeuCure® BNCT system developed by Sumitomo Heavy Industries. This system utilises a beryllium target with a 30 MeV proton beam to generate neutrons. To date, the center has treated over 300 patients and the system is also used for quality control and fundamental research experiments. Information on the stability and longevity of the system is important for facilities and hospitals considering installation of a BNCT system in the future, particularly after an exchange of a major component, such as the target. The beam output (neutron and gamma ray) was measured before and after the exchange of the target. The thermal neutron, fast neutron and gamma ray distribution inside a water phantom was evaluated and the results showed no significant change in the beam quality after the exchange of the target. Furthermore, no significant change in the neutron flux as a function of increasing number of accumulated protons on the target was observed up to a value of just under 400 mAh. The NeuCure® BNCT system has been shown to be highly stable and the frequency of target replacement was much less than lithium target-based accelerators. • BNCT for recurrent head and cancer has been approved in Japan since June 2020 • The BNCT system by Sumitomo Heavy Industries is equipped with a beryllium target • Stability and longevity of the system are important for clinical use • No decrease in the beam output was observed with the beryllium target system • Beryllium target system has better longevity than lithium target systems
Authors
- Naonori Hu
- Ryo Kakino
- Nishiki Matsubayashi (ORCID: https://orcid.org/0000-0002-0090-7759)
- Satoshi Takeno (ORCID: https://orcid.org/0000-0002-8761-3461)
- Keiji Nihei (ORCID: https://orcid.org/0000-0002-9603-0567)
- Mai Nojiri
- Hiroki Tanaka (ORCID: https://orcid.org/0000-0002-6195-3177)
- Minoru Suzuki (ORCID: https://orcid.org/0000-0002-5421-9417)
- Teruhito Aihara
- Kazuhiko Akita
- Koji Ono (ORCID: https://orcid.org/0000-0003-2268-0804)
- Takushi Takata (ORCID: https://orcid.org/0000-0001-6436-6130)
- Yuki Yoshino
- Akinori Sasaki
- Syuushi Yoshikawa
Institutions
- Kansai Medical University (JP)
- Kyoto University (JP)
- Kyoto Prefectural University of Medicine (JP)
- Osaka University Hospital (JP)
- Osaka University of Pharmaceutical Sciences (JP)
Publication Details
- Journal
- Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
- Published
- 2024-12-03
- DOI
- https://doi.org/10.1016/j.nima.2024.170126
- Citations
- 3
- Primary Topic
- Boron Compounds in Chemistry
- Type
- article
- Field-Weighted Citation Impact
- 1.21