Clinical evaluation of performance, stability, and longevity of an accelerator system designed for boron neutron capture therapy utilising a beryllium target

Boron neutron capture therapy for recurrent head and neck cancer has been approved as an insurance covered treatment in Japan since June 2020. The Kansai BNCT Medical Center has installed the NeuCure® BNCT system developed by Sumitomo Heavy Industries. This system utilises a beryllium target with a 30 MeV proton beam to generate neutrons. To date, the center has treated over 300 patients and the system is also used for quality control and fundamental research experiments. Information on the stability and longevity of the system is important for facilities and hospitals considering installation of a BNCT system in the future, particularly after an exchange of a major component, such as the target. The beam output (neutron and gamma ray) was measured before and after the exchange of the target. The thermal neutron, fast neutron and gamma ray distribution inside a water phantom was evaluated and the results showed no significant change in the beam quality after the exchange of the target. Furthermore, no significant change in the neutron flux as a function of increasing number of accumulated protons on the target was observed up to a value of just under 400 mAh. The NeuCure® BNCT system has been shown to be highly stable and the frequency of target replacement was much less than lithium target-based accelerators. • BNCT for recurrent head and cancer has been approved in Japan since June 2020 • The BNCT system by Sumitomo Heavy Industries is equipped with a beryllium target • Stability and longevity of the system are important for clinical use • No decrease in the beam output was observed with the beryllium target system • Beryllium target system has better longevity than lithium target systems

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2024-12-03
DOI
https://doi.org/10.1016/j.nima.2024.170126
Citations
3
Primary Topic
Boron Compounds in Chemistry
Type
article
Field-Weighted Citation Impact
1.21
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical evaluation of performance, stability, and longevity of an accelerator system designed for boron neutron capture therapy utilising a beryllium target

Naonori Hu, Ryo Kakino, Nishiki Matsubayashi, Satoshi Takeno et al.
3 citations
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Boron Compounds in Chemistry
1.21
article

Clinical evaluation of performance, stability, and longevity of an accelerator system designed for boron neutron capture therapy utilising a beryllium target

Naonori Hu, Ryo Kakino, Nishiki Matsubayashi, Satoshi Takeno, Keiji Nihei, Mai Nojiri, Hiroki Tanaka, Minoru Suzuki, Teruhito Aihara, Kazuhiko Akita, Koji Ono, Takushi Takata, Yuki Yoshino, Akinori Sasaki, Syuushi Yoshikawa
article en
3 citations

Abstract

Boron neutron capture therapy for recurrent head and neck cancer has been approved as an insurance covered treatment in Japan since June 2020. The Kansai BNCT Medical Center has installed the NeuCure® BNCT system developed by Sumitomo Heavy Industries. This system utilises a beryllium target with a 30 MeV proton beam to generate neutrons. To date, the center has treated over 300 patients and the system is also used for quality control and fundamental research experiments. Information on the stability and longevity of the system is important for facilities and hospitals considering installation of a BNCT system in the future, particularly after an exchange of a major component, such as the target. The beam output (neutron and gamma ray) was measured before and after the exchange of the target. The thermal neutron, fast neutron and gamma ray distribution inside a water phantom was evaluated and the results showed no significant change in the beam quality after the exchange of the target. Furthermore, no significant change in the neutron flux as a function of increasing number of accumulated protons on the target was observed up to a value of just under 400 mAh. The NeuCure® BNCT system has been shown to be highly stable and the frequency of target replacement was much less than lithium target-based accelerators. • BNCT for recurrent head and cancer has been approved in Japan since June 2020 • The BNCT system by Sumitomo Heavy Industries is equipped with a beryllium target • Stability and longevity of the system are important for clinical use • No decrease in the beam output was observed with the beryllium target system • Beryllium target system has better longevity than lithium target systems

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1072
Kansai Medical University (JP), Kyoto University (JP), Kyoto Prefectural University of Medicine (JP), Osaka University Hospital (JP), Osaka University of Pharmaceutical Sciences (JP)
Affordable and clean energy
Openalex Percentile: Top 19%
Boron Compounds in Chemistry
1.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.