Arrhythmia Detection in Atrioventricular, Single-Lead, Floating Atrial Dipole ICD Systems Compared with Conventional Single- and Dual-Chamber Defibrillators
VDD devices revealed an advantage in atrial arrhythmia detection compared to VVI ICDs and were non-inferior to DDD systems. Their main indication may be closer monitoring in high-risk patients with atrial arrhythmias to help therapy optimization and not the improvement of tachycardia discrimination.
Authors
- László Sághy (ORCID: https://orcid.org/0000-0003-2917-0583)
- Tamás Szili‐Törok (ORCID: https://orcid.org/0000-0002-7165-8243)
- Róbert Pap (ORCID: https://orcid.org/0000-0002-7009-5063)
- M Miklos
- Attila Makai
- A Benak (ORCID: https://orcid.org/0000-0002-6143-8933)
- Peter Gedeon
- Rita B. Gagyi (ORCID: https://orcid.org/0000-0003-4703-2897)
- Máté Vámos (ORCID: https://orcid.org/0000-0003-1611-4443)
- Flóra Diána Gausz (ORCID: https://orcid.org/0000-0002-4058-7710)
- Gábor Bencsik (ORCID: https://orcid.org/0009-0009-0506-8366)
- Kom Nangob Manuela Lena
- Dóra Krányák (ORCID: https://orcid.org/0009-0002-0040-8359)
Institutions
- University of Szeged (HU)
- Queen Elizabeth University Hospital (GB)
Publication Details
- Journal
- Journal of Cardiovascular Development and Disease
- Published
- 2024-12-01
- DOI
- https://doi.org/10.3390/jcdd11120386
- Citations
- 2
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.80