Reconstruction errors in clinical intraoperative TRUS–based prostate HDR-BT detected using electromagnetic tracking
This study found that a quarter of needle reconstructions contained errors >2mm, and that high and severe errors were not uncommon. EMT can play an important role in detecting and preventing these reconstruction errors without disrupting the clinical workflow.
Authors
- Jérémy Godart (ORCID: https://orcid.org/0000-0001-5070-8128)
- Remi A. Nout (ORCID: https://orcid.org/0000-0001-8011-2982)
- Lorne Luthart
- Mischa S. Hoogeman (ORCID: https://orcid.org/0000-0002-4264-9903)
- Ioannis Androulakis (ORCID: https://orcid.org/0000-0003-4816-7048)
- Miranda E.M.C. Christianen
- Henrike Westerveld
- Inger‐Karine K. Kolkman‐Deurloo
Institutions
- Erasmus MC Cancer Institute (NL)
Publication Details
- Journal
- Brachytherapy
- Published
- 2024-12-01
- DOI
- https://doi.org/10.1016/j.brachy.2024.11.004
- Citations
- 5
- Primary Topic
- Prostate Cancer Diagnosis and Treatment
- Type
- article
- Field-Weighted Citation Impact
- 2.09
Funders
- Elekta
- Elekta