The effect of antiseptic solutions on the accuracy in blood glucose measurement using a personal glucose meter

Introduction . Patients with diabetes mellitus perform blood glucose measurements daily using their personal glucose meters. The procedure of disinfecting the place of blood sampling with antiseptic solutions or washing hands with soap and drying them thoroughly is recommended to the patients to prevent infections. However, the effects of used antiseptics on the accuracy in blood glucose measurement using glucose meters have not been evaluated. Aim. To evaluate the effects of antiseptic solutions on the accuracy in blood glucose measurement using a personal glucose meter. Materials and methods. We mixed several drops of the control solution (CS) with low “L” or high “H” concentrations, which were developed for the Contour Plus glucose meter, and a drop of one of five antiseptic solutions: chlorhexidine bigluconate (Chlorhexidine), benzyl dimethyl[3-(myristoylamine)-propyl] ammonium chloride monohydrate (Miramistin), salts of fatty acids (liquid soap), ethyl alcohol, hydrogen peroxide. Thereafter, the blood glucose levels were measured with a glucose meter. The results obtained were compared with the reference range for the CS with appropriate concentration. Results. Chlorhexidine solution had no effect on the blood glucose measurements in the CS. Adding Miramistin solution (1/5 of the sample with CS “L” and “H”) and liquid soap (4/5 of the sample with CS “L” and 2/5 of the sample with CS “H”) resulted in underestimating blood glucose measurements. Alcohol solution had no effect on the blood glucose measurements in the CS “L”, but the glucometer failed to measure the blood glucose concentration in half of the cases, requiring the test strip to be replaced (“E11” error); when mixed with CS “H”, the measurements were overestimated in 2/5 of samples. Adding hydrogen peroxide solution (4/5 of the sample with CR “L” and “H”) resulted in overestimating the blood glucose measurements. Discussion. Getting fruit juice on the hands, applying hydroquinone-containing cosmetics to the skin affects the accuracy in glucose measurements. Correct use of alcohol-containing disinfectants does not affect results of the blood glucose measurement using a personal glucose meter. Conclusion . The results of the blood glucose measurement using a glucose meter can be significantly distorted if antiseptic gets into the sample. Traces of antiseptic should be removed before obtaining the sample that will be used to measure the blood glucose level with a glucose meter. Hydrogen peroxide should not be used as antiseptic solution when using a glucose meter.

Authors

Institutions

Publication Details

Journal
Meditsinskiy sovet = Medical Council
Published
2024-11-26
DOI
https://doi.org/10.21518/ms2024-431
Citations
1
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.66
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect of antiseptic solutions on the accuracy in blood glucose measurement using a personal glucose meter

A. V. Vitebskaya, K. P. Telegina
1 citations
Meditsinskiy sovet = Medical Council
Spectroscopy Techniques in Biomedical and Chemical Research
0.66
article

The effect of antiseptic solutions on the accuracy in blood glucose measurement using a personal glucose meter

A. V. Vitebskaya, K. P. Telegina
article en
1 citations

Abstract

Introduction . Patients with diabetes mellitus perform blood glucose measurements daily using their personal glucose meters. The procedure of disinfecting the place of blood sampling with antiseptic solutions or washing hands with soap and drying them thoroughly is recommended to the patients to prevent infections. However, the effects of used antiseptics on the accuracy in blood glucose measurement using glucose meters have not been evaluated. Aim. To evaluate the effects of antiseptic solutions on the accuracy in blood glucose measurement using a personal glucose meter. Materials and methods. We mixed several drops of the control solution (CS) with low “L” or high “H” concentrations, which were developed for the Contour Plus glucose meter, and a drop of one of five antiseptic solutions: chlorhexidine bigluconate (Chlorhexidine), benzyl dimethyl[3-(myristoylamine)-propyl] ammonium chloride monohydrate (Miramistin), salts of fatty acids (liquid soap), ethyl alcohol, hydrogen peroxide. Thereafter, the blood glucose levels were measured with a glucose meter. The results obtained were compared with the reference range for the CS with appropriate concentration. Results. Chlorhexidine solution had no effect on the blood glucose measurements in the CS. Adding Miramistin solution (1/5 of the sample with CS “L” and “H”) and liquid soap (4/5 of the sample with CS “L” and 2/5 of the sample with CS “H”) resulted in underestimating blood glucose measurements. Alcohol solution had no effect on the blood glucose measurements in the CS “L”, but the glucometer failed to measure the blood glucose concentration in half of the cases, requiring the test strip to be replaced (“E11” error); when mixed with CS “H”, the measurements were overestimated in 2/5 of samples. Adding hydrogen peroxide solution (4/5 of the sample with CR “L” and “H”) resulted in overestimating the blood glucose measurements. Discussion. Getting fruit juice on the hands, applying hydroquinone-containing cosmetics to the skin affects the accuracy in glucose measurements. Correct use of alcohol-containing disinfectants does not affect results of the blood glucose measurement using a personal glucose meter. Conclusion . The results of the blood glucose measurement using a glucose meter can be significantly distorted if antiseptic gets into the sample. Traces of antiseptic should be removed before obtaining the sample that will be used to measure the blood glucose level with a glucose meter. Hydrogen peroxide should not be used as antiseptic solution when using a glucose meter.

Meditsinskiy sovet = Medical Council(19)
D. Mendeleyev University of Chemical Technology of Russia (RU), Sechenov University (RU)
Openalex Percentile: Top 18%
Spectroscopy Techniques in Biomedical and Chemical Research
0.66
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.