Pedicle Screw Pull-Out Strength Between the Jamshidi Needle and Pedicle Probe Techniques in an Osteoporotic Cancellous Bone Model: A Biomechanical Study

Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques. The research sought to determine if the surgical device used in aiding pedicle screw placement, pedicle probe (open technique) and Jamshidi needle (minimally invasive surgical (MIS) technique), affects the eventual stability of the screw in osteoporotic conditions. The findings of this study could enlighten surgical practices, potentially leading to improved clinical outcomes for patients suffering from osteoporosis-related spinal instability. Materials and methods An in vitro biomechanical comparative study was performed whereby pedicle screws were inserted into a standardized polyurethane foam model of grade 10, mimicking osteoporotic bone. Cylindrical poly-axial pedicle screws of 6.5 mm diameter and 45 mm length made out of medical-grade titanium alloy, Ti-6Al-4V, were inserted using four different techniques: Jamshidi needle, Jamshidi needle with tapping, pedicle probe, and pedicle probe with tapping. The screws were inserted in a standardized manner across all groups; the constructs were subsequently attached to the Material Testing System (MTS) 810 machine (MTS Systems Corporation, Eden Prairie, Minnesota, United States) using a customized jig. A direct-load-to-fail test was performed, where data was collected and tabulated into a force-displacement graph. The axial pull-out strength, axial stiffness, and displacement to failure of each construct were then extracted from the graph. Independent samples t-test was then used to compare and study the association between the groups. Results The pedicle probe technique demonstrated superior pull-out strength (698.36±16.34 N) compared to the Jamshidi needle technique (557.15±4.52N) (p<0.05). A greater displacement to failure was also seen in the pedicle probe group (2.26±0.04 mm) versus the Jamshidi needle group (1.18±0.06 mm) (p<0.05). However, the Jamshidi needle technique exhibited higher axial stiffness (336.88±23.24 N/mm) compared to the pedicle probe technique (208.82±7.82 N/mm) (p<0.05). In examining the influence of tapping on both techniques, results show significantly reduced pull-out strength and displacement to failure in the pedicle probe group. Conclusions The pedicle probe technique offers enhanced initial stability in the osteoporotic bone as evidenced by the superior pull-out strength and displacement to failure. On the other hand, the Jamshidi needle technique provides greater resistance to deformation, demonstrated by higher axial stiffness. Tapping should be carefully considered, especially while using the pedicle probe technique, as demonstrated by significantly reduced pull-out strength and displacement to failure. The choice of technique should be informed by specific clinical context balancing the need for initial stability, resistance to deformation, and risk of screw failure.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2024-11-29
DOI
https://doi.org/10.7759/cureus.74747
Citations
2
Primary Topic
Spinal Fractures and Fixation Techniques
Type
article
Field-Weighted Citation Impact
1.15
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pedicle Screw Pull-Out Strength Between the Jamshidi Needle and Pedicle Probe Techniques in an Osteoporotic Cancellous Bone Model: A Biomechanical Study

Fadzrul Abbas Mohamed Ramlee, Mohd Hezery Harun, Shival Tharmaseelan, M. Abdullah
2 citations
Cureus
Spinal Fractures and Fixation Techniques
1.15
article

Pedicle Screw Pull-Out Strength Between the Jamshidi Needle and Pedicle Probe Techniques in an Osteoporotic Cancellous Bone Model: A Biomechanical Study

Fadzrul Abbas Mohamed Ramlee, Mohd Hezery Harun, Shival Tharmaseelan, M. Abdullah
article en
2 citations

Abstract

Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques. The research sought to determine if the surgical device used in aiding pedicle screw placement, pedicle probe (open technique) and Jamshidi needle (minimally invasive surgical (MIS) technique), affects the eventual stability of the screw in osteoporotic conditions. The findings of this study could enlighten surgical practices, potentially leading to improved clinical outcomes for patients suffering from osteoporosis-related spinal instability. Materials and methods An in vitro biomechanical comparative study was performed whereby pedicle screws were inserted into a standardized polyurethane foam model of grade 10, mimicking osteoporotic bone. Cylindrical poly-axial pedicle screws of 6.5 mm diameter and 45 mm length made out of medical-grade titanium alloy, Ti-6Al-4V, were inserted using four different techniques: Jamshidi needle, Jamshidi needle with tapping, pedicle probe, and pedicle probe with tapping. The screws were inserted in a standardized manner across all groups; the constructs were subsequently attached to the Material Testing System (MTS) 810 machine (MTS Systems Corporation, Eden Prairie, Minnesota, United States) using a customized jig. A direct-load-to-fail test was performed, where data was collected and tabulated into a force-displacement graph. The axial pull-out strength, axial stiffness, and displacement to failure of each construct were then extracted from the graph. Independent samples t-test was then used to compare and study the association between the groups. Results The pedicle probe technique demonstrated superior pull-out strength (698.36±16.34 N) compared to the Jamshidi needle technique (557.15±4.52N) (p<0.05). A greater displacement to failure was also seen in the pedicle probe group (2.26±0.04 mm) versus the Jamshidi needle group (1.18±0.06 mm) (p<0.05). However, the Jamshidi needle technique exhibited higher axial stiffness (336.88±23.24 N/mm) compared to the pedicle probe technique (208.82±7.82 N/mm) (p<0.05). In examining the influence of tapping on both techniques, results show significantly reduced pull-out strength and displacement to failure in the pedicle probe group. Conclusions The pedicle probe technique offers enhanced initial stability in the osteoporotic bone as evidenced by the superior pull-out strength and displacement to failure. On the other hand, the Jamshidi needle technique provides greater resistance to deformation, demonstrated by higher axial stiffness. Tapping should be carefully considered, especially while using the pedicle probe technique, as demonstrated by significantly reduced pull-out strength and displacement to failure. The choice of technique should be informed by specific clinical context balancing the need for initial stability, resistance to deformation, and risk of screw failure.

CureusVol. 16(11)
Universiti Putra Malaysia (MY), University of Kuala Lumpur (MY)
Openalex Percentile: Top 19%
Spinal Fractures and Fixation Techniques
1.15
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.