Fluence smoothing evaluation for whole‐breast automatically generated treatment plans
Our findings suggest that using a ratio of smoothing weights to target priorities between 1:1.5 and 1:1.6 leads to a favorable balance between complexity and dosimetric plan quality, with no significant impacts on dose constraint adherence.