This work presents the design, fabrication, and rigorous validation of a flexible, wireless, capacitive pressure sensor for the full-range continuous monitoring of ventricular pressure. The proposed system consists of an implantable set and an external readout device; both modules were designed to form an RCL resonant circuit for passive, wireless pressure sensing and signal retrieving. Using surface micromachining and flexible electronics techniques, a two-variable capacitor array and a dual-layer planar coil were integrated into a flexible ergonomic substrate, avoiding hybrid-like connections in the implantable set. The proposed arrangement (capacitor array and dual-layer coil) allows us to optimize the operation pressure range and sensing distance. The use of polyimide as both the flexible substrate and the passivation material is a key feature, ensuring a biocompatible, implantable set that is mechanically flexible and can be folded to a compact size to achieve minimally invasive implantation. An external readout device has also been developed using a discrete printed circuit board (PCB) approach to support pressure measurements. The pressure responsivity of the sensor was validated to the laboratory level using a controlled pressure chamber. The results obtained show that the capacitance value of the sensor changed from 5.68 pF to 33.26 pF as the pressure varied from 0 to 300 mmHg. Correspondingly, the resonance frequency of the implantable set shifted from 12.75 MHz to 5.27 MHz. The sensitivity of the capacitive sensor was approximately 0.58 pF/mmHg and the typical response time was 220 ms. The wireless system performance was evaluated in both air and synthetic biological tissue using a Maxwell-Wien bridge circuit. The results showed a sensing distance longer than 3.5 cm, even under moderate misalignment conditions (up to 1.5 cm). The output voltage was successfully measured, ranging from 502.54 mV to 538.29 mV, throughout the full pressure range, with a measurement error of ±2.2 mV.

Authors

Institutions

Publication Details

Journal
Micromachines
Published
2024-11-28
DOI
https://doi.org/10.3390/mi15121435
Citations
3
Primary Topic
Wireless Power Transfer Systems
Type
article
Field-Weighted Citation Impact
0.63
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fabrication of an Integrated, Flexible, Wireless Pressure Sensor Array for the Monitoring of Ventricular Pressure

W. Calleja-Arriaga, Natiely Hernández-Sebastián, Bernardino Barrientos-García, Daniela Díaz-Alonso et al.
3 citations
Micromachines
Wireless Power Transfer Systems
0.63
article

Fabrication of an Integrated, Flexible, Wireless Pressure Sensor Array for the Monitoring of Ventricular Pressure

W. Calleja-Arriaga, Natiely Hernández-Sebastián, Bernardino Barrientos-García, Daniela Díaz-Alonso, Francisco J. Renero-Carrillo
article en
3 citations

Abstract

This work presents the design, fabrication, and rigorous validation of a flexible, wireless, capacitive pressure sensor for the full-range continuous monitoring of ventricular pressure. The proposed system consists of an implantable set and an external readout device; both modules were designed to form an RCL resonant circuit for passive, wireless pressure sensing and signal retrieving. Using surface micromachining and flexible electronics techniques, a two-variable capacitor array and a dual-layer planar coil were integrated into a flexible ergonomic substrate, avoiding hybrid-like connections in the implantable set. The proposed arrangement (capacitor array and dual-layer coil) allows us to optimize the operation pressure range and sensing distance. The use of polyimide as both the flexible substrate and the passivation material is a key feature, ensuring a biocompatible, implantable set that is mechanically flexible and can be folded to a compact size to achieve minimally invasive implantation. An external readout device has also been developed using a discrete printed circuit board (PCB) approach to support pressure measurements. The pressure responsivity of the sensor was validated to the laboratory level using a controlled pressure chamber. The results obtained show that the capacitance value of the sensor changed from 5.68 pF to 33.26 pF as the pressure varied from 0 to 300 mmHg. Correspondingly, the resonance frequency of the implantable set shifted from 12.75 MHz to 5.27 MHz. The sensitivity of the capacitive sensor was approximately 0.58 pF/mmHg and the typical response time was 220 ms. The wireless system performance was evaluated in both air and synthetic biological tissue using a Maxwell-Wien bridge circuit. The results showed a sensing distance longer than 3.5 cm, even under moderate misalignment conditions (up to 1.5 cm). The output voltage was successfully measured, ranging from 502.54 mV to 538.29 mV, throughout the full pressure range, with a measurement error of ±2.2 mV.

MicromachinesVol. 15(12)
Centro de Investigaciones en Optica (MX), National Institute of Astrophysics, Optics and Electronics (MX)
Affordable and clean energy
Openalex Percentile: Top 30%
Wireless Power Transfer Systems
0.63
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.