Accurate analyses and management of ion concentrations are crucial in precision agriculture. Modern technology-based methods are non-destructive and do not require sample preparation, enabling fast and accurate analysis; however, they have limitations when processing multiple samples. In this study, a multi-ion analysis system was developed for the prompt and accurate analysis of concentrations of important ions such as NO3−, HnPO4−, K+, Ca2+, and Mg2+. The RGB sensitivity control was automated through calibrations by applying a reference slot-based error rate across six slots between sample measurements, facilitating sample-to-sample comparisons and enabling accurate concentration analysis. By analyzing the correlation between each ion concentration and the proximal absorbance-based concentration prediction in the simultaneous analysis system, the accuracy was verified by achieving a coefficient of determination exceeding 0.99 for most ions. This system minimizes possible deviations between slots by using an automatic calibration algorithm, thereby facilitating the simultaneous analysis of multiple samples. This is important for saving time and cost and can help in real-time nutrient analysis and monitoring in agriculture.

Authors

Institutions

Publication Details

Journal
Agriculture
Published
2024-11-28
DOI
https://doi.org/10.3390/agriculture14122171
Citations
1
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Proximal Absorbance Calibration Method Using an Embedded Blank Reference RGB Sensor for Determination of Ion Concentrations

Yehun Lee, Jung-Kyu Lee, Dong-Hoon Lee
1 citations
Agriculture
Analytical Chemistry and Sensors
0.29
article

Proximal Absorbance Calibration Method Using an Embedded Blank Reference RGB Sensor for Determination of Ion Concentrations

Yehun Lee, Jung-Kyu Lee, Dong-Hoon Lee
article en
1 citations

Abstract

Accurate analyses and management of ion concentrations are crucial in precision agriculture. Modern technology-based methods are non-destructive and do not require sample preparation, enabling fast and accurate analysis; however, they have limitations when processing multiple samples. In this study, a multi-ion analysis system was developed for the prompt and accurate analysis of concentrations of important ions such as NO3−, HnPO4−, K+, Ca2+, and Mg2+. The RGB sensitivity control was automated through calibrations by applying a reference slot-based error rate across six slots between sample measurements, facilitating sample-to-sample comparisons and enabling accurate concentration analysis. By analyzing the correlation between each ion concentration and the proximal absorbance-based concentration prediction in the simultaneous analysis system, the accuracy was verified by achieving a coefficient of determination exceeding 0.99 for most ions. This system minimizes possible deviations between slots by using an automatic calibration algorithm, thereby facilitating the simultaneous analysis of multiple samples. This is important for saving time and cost and can help in real-time nutrient analysis and monitoring in agriculture.

AgricultureVol. 14(12)
Chungbuk National University (KR)
National Research Foundation, National Research Foundation of Korea, Ministry of Education, India
Zero hunger
Openalex Percentile: Top 45%
Analytical Chemistry and Sensors
0.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Proximal Absorbance Calibration Method Using an Embedded Blank Reference RGB Sensor for Determination of Ion Concentrations — Yehun Lee, Jung-Kyu Lee, et al. · Agriculture (2024) | TGRS Research Map | TGRS