Performance evaluation of enzymatic total bile acid (TBA) routine assays: systematic comparison of five fifth-generation TBA cycling methods and their individual bile acid recovery from HPLC-MS/MS reference
The linear relationship and constant proportional bias between all five routine assays enable the harmonization of TBA measurements up to 60 μmol/L. However, for patient samples with high TBA levels and disease-specific overrepresentation of individual BAs, harmonization will require: i) optimized reaction conditions to equalize substrate specificity, and ii) calibration to a common, commutable reference material with well-defined BA composition instead of internal standards spiked with different BAs.