Development of dedicated phantom for end‐to‐end testing of multiple image‐guidance configurations incorporating light‐section‐based surface guidance system
The new E2E phantom can be used to evaluate the detection accuracy of a light-section-based SGRT system and the coordinate coincidence using a Winston-Lutz-based method in multiple image-guided configurations. The detection accuracy in the three different reference images of VOXELAN using this phantom improved to within 1 mm in all directions.
Authors
- Hiroshi Amma
- Yuta Muraki
- Keisuke Yasui (ORCID: https://orcid.org/0000-0002-5581-463X)
- Naoki Hayashi (ORCID: https://orcid.org/0000-0002-1720-3651)
- M Nozue
- K. Ouchi
- Tatsunori Saito
Institutions
- Fujita Health University (JP)
- Seirei Hamamatsu General Hospital (JP)
- Japanese Red Cross Nagoya Daiichi Hospital (JP)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2024-11-26
- DOI
- https://doi.org/10.1002/acm2.14517
- Citations
- 3
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 1.82