Blood-glucose regulator design for diabetics based on LQIR-driven Sliding-Mode-Controller with self-adaptive reaching law

Type I Diabetes is an endocrine disorder that prevents the pancreas from regulating the blood glucose (BG) levels in a patient's body. The ubiquitous Linear-Quadratic-Integral-Regulator (LQIR) is an optimal glycemic regulation strategy; however, it is not resilient enough to withstand measurement noise and meal disruptions. The Sliding-Mode-Controller (SMC) yields robust BG regulation effort at the expense of a discontinuous insulin infusion rate that perturbs the BG concentrations. Hence, the novel contribution of this article is the formulation of a hybridized LQIR-driven SMC strategy that retrieves the benefits of the aforesaid control schemes while avoiding their inherent problems. The proposed control approach is realized by linearly combining a glycemic LQIR law with an innovative sign function sliding mode reaching law that is driven by a customized LQIR-driven sliding surface. The hybridized control scheme generates optimal control decisions yielded by the LQIR while mimicking the robustness characteristic of SMC against bounded exogenous disturbances. Additionally, the SMC reaching law in the proposed control scheme is augmented with a nonlinear adaptation mechanism that flexibly modulates the control activity to effectively compensate for the external perturbations while minimizing the chattering content. The controller parameters are numerically optimized offline. The efficacy of the prescribed hybrid control law is analyzed via customized MATLAB simulations that normalize the patient's BG level to 80 mg/dL, under measurement noise and meal disruptions, from an initial hyperglycemic state. The results justify the improved BG regulation accuracy and disturbance-rejection capability of the proposed control procedure.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2024-11-27
DOI
https://doi.org/10.1371/journal.pone.0314479
Citations
4
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
1.43
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Blood-glucose regulator design for diabetics based on LQIR-driven Sliding-Mode-Controller with self-adaptive reaching law

Omer Saleem, Jamshed Iqbal
4 citations
PLoS ONE
Diabetes Management and Research
1.43
article

Blood-glucose regulator design for diabetics based on LQIR-driven Sliding-Mode-Controller with self-adaptive reaching law

Omer Saleem, Jamshed Iqbal
article en
4 citations

Abstract

Type I Diabetes is an endocrine disorder that prevents the pancreas from regulating the blood glucose (BG) levels in a patient's body. The ubiquitous Linear-Quadratic-Integral-Regulator (LQIR) is an optimal glycemic regulation strategy; however, it is not resilient enough to withstand measurement noise and meal disruptions. The Sliding-Mode-Controller (SMC) yields robust BG regulation effort at the expense of a discontinuous insulin infusion rate that perturbs the BG concentrations. Hence, the novel contribution of this article is the formulation of a hybridized LQIR-driven SMC strategy that retrieves the benefits of the aforesaid control schemes while avoiding their inherent problems. The proposed control approach is realized by linearly combining a glycemic LQIR law with an innovative sign function sliding mode reaching law that is driven by a customized LQIR-driven sliding surface. The hybridized control scheme generates optimal control decisions yielded by the LQIR while mimicking the robustness characteristic of SMC against bounded exogenous disturbances. Additionally, the SMC reaching law in the proposed control scheme is augmented with a nonlinear adaptation mechanism that flexibly modulates the control activity to effectively compensate for the external perturbations while minimizing the chattering content. The controller parameters are numerically optimized offline. The efficacy of the prescribed hybrid control law is analyzed via customized MATLAB simulations that normalize the patient's BG level to 80 mg/dL, under measurement noise and meal disruptions, from an initial hyperglycemic state. The results justify the improved BG regulation accuracy and disturbance-rejection capability of the proposed control procedure.

PLoS ONEVol. 19(11)
University of Hull (GB), National University of Computer and Emerging Sciences (PK)
Peace, Justice and strong institutions
Openalex Percentile: Top 15%
Diabetes Management and Research
1.43
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.