An Efficient Method to Mitigate RF-induced Heating in DBS patient at 7T Using Geometrically Adjustable RF Head Array.

Motivation: Patients with deep brain stimulation (DBS) devices are strictly prohibited from undergoing 7 Tesla magnetic resonance imaging (MRI) scans. Goal(s): Mitigating radiofrequency (RF)-induced heating in DBS for 7 T MRI using a convenient approach. Approach: By utilizing the electromagnetic characteristics arising from the near-field region of the RF coil, a novel method of rearranging the positions of the RF channels adjacent to the DBS lead using a geometrically adjustable RF array was proposed. Results: Adjusting the position of channels adjacent to lead reduces the specific absorption rate and the extent of temperature increase around the electrode up to 34% and 23.8 %, respectively. Impact: Our method involves strategically adjusting the position of RF coil near the lead, providing a practical solution to mitigate RF-induced heating in DBS. It eliminates the necessity for a new lead design or complex numerical optimization for incident electromagnetic field.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/3744
Primary Topic
Systemic Sclerosis and Related Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Efficient Method to Mitigate RF-induced Heating in DBS patient at 7T Using Geometrically Adjustable RF Head Array.

Youngdae Cho, Hyoungsuk Yoo
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Systemic Sclerosis and Related Diseases
article

An Efficient Method to Mitigate RF-induced Heating in DBS patient at 7T Using Geometrically Adjustable RF Head Array.

Youngdae Cho, Hyoungsuk Yoo
article en

Abstract

Motivation: Patients with deep brain stimulation (DBS) devices are strictly prohibited from undergoing 7 Tesla magnetic resonance imaging (MRI) scans. Goal(s): Mitigating radiofrequency (RF)-induced heating in DBS for 7 T MRI using a convenient approach. Approach: By utilizing the electromagnetic characteristics arising from the near-field region of the RF coil, a novel method of rearranging the positions of the RF channels adjacent to the DBS lead using a geometrically adjustable RF array was proposed. Results: Adjusting the position of channels adjacent to lead reduces the specific absorption rate and the extent of temperature increase around the electrode up to 34% and 23.8 %, respectively. Impact: Our method involves strategically adjusting the position of RF coil near the lead, providing a practical solution to mitigate RF-induced heating in DBS. It eliminates the necessity for a new lead design or complex numerical optimization for incident electromagnetic field.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Affordable and clean energy
Openalex Percentile: Top 65%
Systemic Sclerosis and Related Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies