An Efficient Method to Mitigate RF-induced Heating in DBS patient at 7T Using Geometrically Adjustable RF Head Array.
Motivation: Patients with deep brain stimulation (DBS) devices are strictly prohibited from undergoing 7 Tesla magnetic resonance imaging (MRI) scans. Goal(s): Mitigating radiofrequency (RF)-induced heating in DBS for 7 T MRI using a convenient approach. Approach: By utilizing the electromagnetic characteristics arising from the near-field region of the RF coil, a novel method of rearranging the positions of the RF channels adjacent to the DBS lead using a geometrically adjustable RF array was proposed. Results: Adjusting the position of channels adjacent to lead reduces the specific absorption rate and the extent of temperature increase around the electrode up to 34% and 23.8 %, respectively. Impact: Our method involves strategically adjusting the position of RF coil near the lead, providing a practical solution to mitigate RF-induced heating in DBS. It eliminates the necessity for a new lead design or complex numerical optimization for incident electromagnetic field.
Authors
- Youngdae Cho (ORCID: https://orcid.org/0000-0002-0521-9103)
- Hyoungsuk Yoo
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3744
- Primary Topic
- Systemic Sclerosis and Related Diseases
- Type
- article
- Field-Weighted Citation Impact
- 0.00