Impact of RF imperfections on MRF-based and reference B1+ mapping methods with a commercial 1Tx/32Rx head coil
Motivation: Magnetic Resonance Fingerprinting (MRF) and preparation-based reference B1+ mapping approaches yielded an unexpected B1+ offset. Goal(s): Identify the origin of the B1+ offset. Approach: Pick-up coil measurements to evaluate the actual RF pulse shape using a commercial 1Tx/32Rx head-coil at 7T. Results: For the preparation-based B1+ mapping method with rectangular pulses, longer RF rise times (~100 µs) and phase variations (~10°) were observed. Furthermore, during the 21s-long MRF acquisition using sinc-pulses, an increased RF amplitude (~2.5%) was observed. By incorporating the measured RF alterations, the B1+ offset could be substantially reduced. Impact: Unidentified effects of the RF transmission such as deviation in envelope and temporal instability affect measurements. Characterizing the RF transmission enables quantification and correction of these deviations.
Authors
- Stephan Orzada (ORCID: https://orcid.org/0000-0001-9784-4354)
- Sebastian Schmitter (ORCID: https://orcid.org/0000-0003-4410-6790)
- Berk Silemek (ORCID: https://orcid.org/0000-0001-8227-3632)
- Max Lutz (ORCID: https://orcid.org/0009-0001-1956-3757)
- Lance DelaBarre
- Christoph Stefan Aigner (ORCID: https://orcid.org/0000-0003-3618-9610)
- Frank Seifert (ORCID: https://orcid.org/0000-0002-7065-2528)
- Tobias Schaeffter
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0473
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00