Motion-Corrected Subspace Navigator Reconstruction for High Resolution Spiral First-pass Myocardial Perfusion Imaging at 3 Tesla

Motivation: As perfusion images are typically acquired over 60 heart beats, respiratory motion is unavoidable. Motion compromises spatio-temporal reconstructions. Goal(s): Residual undersampling artifacts and contrast variation make deformation field estimation challenging. This work aims to get an accurate deformation field from the auxiliary reconstruction and incorporate it into the forward reconstruction model to improve perfusion images. Approach: To obtain high-quality images for motion estimation, the fixed-angle spiral navigator is used to extract temporal basis. The rigid and non-rigid motion corrections are jointly incorporated into the subspace reconstruction. Results: Motion-corrected whole-heart first-pass spiral myocardial perfusion imaging with a high resolution of 1.3 mm2 is achieved. Impact: The proposed navigator-guided subspace motion correction reconstruction pipeline substantially improves the image quality, sharpness, and alignment of the 1.3mm² high-resolution spiral myocardial perfusion imaging, benefiting voxel-wise perfusion quantification crucial for assessing ischemic heart disease.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/0271
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Motion-Corrected Subspace Navigator Reconstruction for High Resolution Spiral First-pass Myocardial Perfusion Imaging at 3 Tesla

Junyu Wang, Quan Chen, Sizhuo Liu, Michael Salerno et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Motion-Corrected Subspace Navigator Reconstruction for High Resolution Spiral First-pass Myocardial Perfusion Imaging at 3 Tesla

Junyu Wang, Quan Chen, Sizhuo Liu, Michael Salerno, Shen Zhao, Xitong Wang
article en

Abstract

Motivation: As perfusion images are typically acquired over 60 heart beats, respiratory motion is unavoidable. Motion compromises spatio-temporal reconstructions. Goal(s): Residual undersampling artifacts and contrast variation make deformation field estimation challenging. This work aims to get an accurate deformation field from the auxiliary reconstruction and incorporate it into the forward reconstruction model to improve perfusion images. Approach: To obtain high-quality images for motion estimation, the fixed-angle spiral navigator is used to extract temporal basis. The rigid and non-rigid motion corrections are jointly incorporated into the subspace reconstruction. Results: Motion-corrected whole-heart first-pass spiral myocardial perfusion imaging with a high resolution of 1.3 mm2 is achieved. Impact: The proposed navigator-guided subspace motion correction reconstruction pipeline substantially improves the image quality, sharpness, and alignment of the 1.3mm² high-resolution spiral myocardial perfusion imaging, benefiting voxel-wise perfusion quantification crucial for assessing ischemic heart disease.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Good health and well-being
Openalex Percentile: Top 62%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Motion-Corrected Subspace Navigator Reconstruction for High Resolution Spiral First-pass Myocardial Perfusion Imaging at 3 Tesla — Junyu Wang, Quan Chen, et al. · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2024) | TGRS Research Map | TGRS