Accuracy of Measuring Opening and Closing Characteristics of the Aortic Valve with SPEEDI MRI
Motivation: get-SPEEDI, a recently published pulse sequence, offers a promising MRI technique for imaging the rapid dynamics of the aortic valve (AV) opening and closing with sub-millisecond temporal resolution. Goal(s): This study aims to assess the accuracy of measuring the AV dynamics with get-SPEEDI by comparing its performance with transthoracic ultrasound, which is the gold standard clinically. Approach: Ultrasound echocardiograms and get-SPEEDI MR images of AV were acquired in healthy human subjects with 0.6- and 0.8-ms temporal resolution, respectively. Results: There were no statistically significant differences between get-SPEEDI and ultrasound in the measurements of AV opening and closing dynamics and the maximum AV area. Impact: The dynamic characteristics of the aortic valve measured with get-SPEEDI MRI agree well with the ultrasound measurements. get-SPEEDI MRI provides a new imaging tool for diagnosis of aortic valve diseases.
Authors
- Alessandro Scotti (ORCID: https://orcid.org/0000-0002-8892-881X)
- Qingfei Luo (ORCID: https://orcid.org/0000-0001-5574-817X)
- Xiaohong Zhou
- Yosman Dhar
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0280
- Primary Topic
- Cardiac Valve Diseases and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00