EPI Phase Correction revisited - Beat Phenomena and Transient States

Motivation: For EPI, beat phenomena can occur and result in a slowly varying, erroneous phase component caused by mechanical resonances of the gradient system. Current phase correction approaches do not account for this error. Goal(s): Improving EPI correction by characterization of gradient-induced, time-varying errors of zeroth and first spatial order. Approach: EPI image projections of a spherical phantom were acquired by omitting the phase encoding blips. The resulting x-ky-data enable a characterization of zeroth- and first-order effects separately. Results: The mechanical properties of the gradient system result in a transient state during which an erroneous, temporally and spatially modulated magnetic field gradient occurs. Impact: EPI is a widely applied fast imaging technique but sensitive to gradient-induced magnetic field deviations. Therefore, phase correction is of utmost importance and may be improved by the proposed characterization of slowly varying, erroneous phases.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/3942
Primary Topic
Particle accelerators and beam dynamics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

EPI Phase Correction revisited - Beat Phenomena and Transient States

Sebastian Kozerke, Tobias Schaeffter, Hannes Dillinger
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Particle accelerators and beam dynamics
article

EPI Phase Correction revisited - Beat Phenomena and Transient States

Sebastian Kozerke, Tobias Schaeffter, Hannes Dillinger
article en

Abstract

Motivation: For EPI, beat phenomena can occur and result in a slowly varying, erroneous phase component caused by mechanical resonances of the gradient system. Current phase correction approaches do not account for this error. Goal(s): Improving EPI correction by characterization of gradient-induced, time-varying errors of zeroth and first spatial order. Approach: EPI image projections of a spherical phantom were acquired by omitting the phase encoding blips. The resulting x-ky-data enable a characterization of zeroth- and first-order effects separately. Results: The mechanical properties of the gradient system result in a transient state during which an erroneous, temporally and spatially modulated magnetic field gradient occurs. Impact: EPI is a widely applied fast imaging technique but sensitive to gradient-induced magnetic field deviations. Therefore, phase correction is of utmost importance and may be improved by the proposed characterization of slowly varying, erroneous phases.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 66%
Particle accelerators and beam dynamics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.