Rapid, motion robust T2w and T2FLAIR imaging
Motivation: The T2w and T2FLAIR are extremely versatile MRI contrasts and paramount to asses a wide range of pathologies in neuroimaging. Goal(s): A highly efficient acquisition strategy is proposed that jointly acquires T2w and T2FLAIR full brain data in a little more than half a minute. Approach: Single shot FSE readouts, inversion pulses and saturation bands are strategically interleaved to maximize sampling efficiency and avoid dummy TRs. Prospective motion correction is added to increase robustness. Results: Volunteer results comparing different protocols, a motion experiment and a pediatric scan are presented. Acceptable image quality was achieved even under severe motion. Impact: A half minute, motion robust, whole brain T2w and T2FLAIR scan is very interesting for clinical neuroimaging. It is potentially very useful for acute MR, pediatric imaging, screening, disease progression or treatment monitoring.
Authors
- Sophie Schauman (ORCID: https://orcid.org/0000-0002-3744-2553)
- Adam van Niekerk (ORCID: https://orcid.org/0000-0001-9731-6930)
- Tim Sprenger (ORCID: https://orcid.org/0000-0001-7037-8605)
- Ola Norbeck (ORCID: https://orcid.org/0000-0002-0518-5206)
- Henric Rydén (ORCID: https://orcid.org/0000-0003-3269-7482)
- Stefan Skare (ORCID: https://orcid.org/0000-0001-5403-2153)
- Enrico Aventi
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3252
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00