Instrument Power Monitoring Analysis of Commercial Point-of-Care MR in Resource-Constrained Healthcare Settings: Initial Feasibility

Motivation: The 64mT Hyperfine Swoop (Hyperfine Inc. Guilford CT) is a first-of-kind point-of-care (POC) commercial system to allow ‘patient-to-scanner’ imaging. Goal(s): We report an independent technical benchmark analysis of this system’s power consumption in this study (pass-through measures of W, kWh, and A). Approach: Performance benchmarks of MR systems power consumption was measured per-pulse sequence using consumer-grade instrumentation under continuous monitoring. Custom acquisition protocol using two portable 120VAC NEMA-standard power banks were examined. Results: A look-up benchmark table of empirical surge current draw (with 120VAC), its implicit advisory, and a potentially viable protocol run example without 'wall' i.e. via stand-alone power supply are reported. Impact: POC-MR Instrumentation power benchmark considerations on a per-pulse-sequence power consumption basis provide key insights into protocol deployment, scheduling, and optimal scan resource management considerations in resource-limited settings. Successful pulse sequence protocol implementation with under resource-limited setting was also demonstrated.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/2840
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Instrument Power Monitoring Analysis of Commercial Point-of-Care MR in Resource-Constrained Healthcare Settings: Initial Feasibility

Keigo Kawaji, John G. Georgiadis, Liam Crosby, Sudarshan Ragunathan et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Instrument Power Monitoring Analysis of Commercial Point-of-Care MR in Resource-Constrained Healthcare Settings: Initial Feasibility

Keigo Kawaji, John G. Georgiadis, Liam Crosby, Sudarshan Ragunathan, Vivian Nguyen, S. S. Sandhu, Natalie Hamill, Dave Tailor
article en

Abstract

Motivation: The 64mT Hyperfine Swoop (Hyperfine Inc. Guilford CT) is a first-of-kind point-of-care (POC) commercial system to allow ‘patient-to-scanner’ imaging. Goal(s): We report an independent technical benchmark analysis of this system’s power consumption in this study (pass-through measures of W, kWh, and A). Approach: Performance benchmarks of MR systems power consumption was measured per-pulse sequence using consumer-grade instrumentation under continuous monitoring. Custom acquisition protocol using two portable 120VAC NEMA-standard power banks were examined. Results: A look-up benchmark table of empirical surge current draw (with 120VAC), its implicit advisory, and a potentially viable protocol run example without 'wall' i.e. via stand-alone power supply are reported. Impact: POC-MR Instrumentation power benchmark considerations on a per-pulse-sequence power consumption basis provide key insights into protocol deployment, scheduling, and optimal scan resource management considerations in resource-limited settings. Successful pulse sequence protocol implementation with under resource-limited setting was also demonstrated.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 62%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.