Zoomed imaging with calibration-free destructive B1+ shimming for rapid fMRI
Motivation: High-resolution MRI enables precise studies of specific brain regions but this comes at a cost of increased acquisition time. Goal(s): We examined if calibration-free destructive interferences can be introduced with simple static B1-shims for zoomed fMRI acquisitions. Approach: We optimised destructive B1 phase offsets and implemented these whilst assessing image quality in MPRAGE, tSNR in an EPI timeseries, flip angle distributions and |B1|. Results: Calibration-free destructive shims allowed reducing the field-of-view (50%) and acquisition-time (35%), while retaining good signal (higher B1 and similar tSNR) in the target (visual cortex). Impact: The calibration-free destructive B1 shims can be applied across brain regions with standard head coils and therefore translate across neuroscientific studies, while reducing acquisition-time and signal contributions from noise-prone areas. This may flexibly improve the spatial resolution and signal quality.
Authors
- Wietske van der Zwaag (ORCID: https://orcid.org/0000-0003-3223-9721)
- Nikos Priovoulos (ORCID: https://orcid.org/0000-0001-5661-6670)
- Emma Brouwer (ORCID: https://orcid.org/0000-0003-1355-8057)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3301
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00