Real-Time Imaging of Lower Airway Collapse at 0.55T

Motivation: Tracheomalacia involves intermittent collapse of the lower airway and is difficult to diagnose. Currently, this condition is evaluated with invasive procedures like bronchoscopy or dynamic computer tomography (CT) which utilizes ionizing radiation. Goal(s): To develop and optimize a high spatial and temporal resolution pulse sequence for evaluating tracheomalacia at 0.55T. Approach: We optimize contrast (sequence, flip angle) for dynamic imaging in the lower airway for both 2D and SMS real-time imaging. Results: Spiral acquisition achieves sufficient temporal resolution and the TrueFISP sequence (flip angle=32°) offers optimal contrast for imaging lower airway collapse. Impact: We demonstrate the feasibility of capturing lower airway collapse with 2D and SMS real-time imaging (at 0.55T) in a way that is informative to the diagnosis and longitudinal monitoring of tracheomalacia.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/0298
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-Time Imaging of Lower Airway Collapse at 0.55T

Roberta Miyeko Kato, Ecrin Yağız, Krishna S. Nayak, Adrienne E. Campbell‐Washburn et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Airway Management and Intubation Techniques
article

Real-Time Imaging of Lower Airway Collapse at 0.55T

Roberta Miyeko Kato, Ecrin Yağız, Krishna S. Nayak, Adrienne E. Campbell‐Washburn, Marcus Y. Chen, Marcela A. Ferrada, Sarina Kapai, Ye Tian, Prakash Kumar
article en

Abstract

Motivation: Tracheomalacia involves intermittent collapse of the lower airway and is difficult to diagnose. Currently, this condition is evaluated with invasive procedures like bronchoscopy or dynamic computer tomography (CT) which utilizes ionizing radiation. Goal(s): To develop and optimize a high spatial and temporal resolution pulse sequence for evaluating tracheomalacia at 0.55T. Approach: We optimize contrast (sequence, flip angle) for dynamic imaging in the lower airway for both 2D and SMS real-time imaging. Results: Spiral acquisition achieves sufficient temporal resolution and the TrueFISP sequence (flip angle=32°) offers optimal contrast for imaging lower airway collapse. Impact: We demonstrate the feasibility of capturing lower airway collapse with 2D and SMS real-time imaging (at 0.55T) in a way that is informative to the diagnosis and longitudinal monitoring of tracheomalacia.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 60%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.