In-vivo Magnetization Transfer neuroimaging at 46 mT
Motivation: Here, we exploited the Low SAR in low-field MRI to study magnetization transfer (MT) , and potential to enhance brain tissue contrast, in particular between white and gray matter (challenging at low-field). Goal(s): The study aimed to identify suitable MT settings and explore MT contrast in-vivo at 46 mT. Approach: MT settings were optimized to minimize direct saturation and maximize MT contrast. With this setup, we scanned five volunteers at 46 mT and acquired additional 3T scans for two participantsResults: Optimal MT settings were defined in the phantom and in-vivo scans demonstrated the expected MT effects, and improved tissue contrast at 46 mT. Impact: Magnetization transfer could be achieved off-resonance in-vivo at low-field MRI. It improves brain tissue contrast, in particular between white matter and gray matter, and has minimal direct saturation.
Authors
- Beatrice Lena (ORCID: https://orcid.org/0000-0001-9283-5918)
- Thomas O’Reilly (ORCID: https://orcid.org/0000-0002-4547-2955)
- Andrew Webb (ORCID: https://orcid.org/0000-0003-4045-9732)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2825
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00