bSSFP Phase Contrast (PC-SSFP) at 0.55T for Aortic Flow

Motivation: Aortic flow imaging at 0.55T has limited SNR, especially in diastolic phases, which can limit the accuracy of flow and regurgitant fraction measurements. Goal(s): Our goal was to improve the SNR of phase contrast images at 0.55T using a bSSFP readout and maintain accurate flow quantification. Approach: We developed a through-plane phase contrast sequence with bSSFP readout and tested it in phantoms and healthy volunteers. Results: Significantly improved SNR (25.5±9.6 vs 8.2±2.9, p<0.001) and accuracy of velocities measurement were demonstrated in phantom. Accurate flow and constant high SNR were measured using our method in 8 subjects, for both breath-hold and free-breathing cases. Impact: Our proposed phase contrast with bSSFP readout can reliably evaluate aortic flow at 0.55T with higher SNR and VNR leading to more accurate quantification compared to conventional GRE based phase contrast method.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/0281
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

bSSFP Phase Contrast (PC-SSFP) at 0.55T for Aortic Flow

Dana C. Peters, Rajiv Ramasawmy, Adrienne E. Campbell‐Washburn, Jie Xiang et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

bSSFP Phase Contrast (PC-SSFP) at 0.55T for Aortic Flow

Dana C. Peters, Rajiv Ramasawmy, Adrienne E. Campbell‐Washburn, Jie Xiang, Felicia Seemann
article en

Abstract

Motivation: Aortic flow imaging at 0.55T has limited SNR, especially in diastolic phases, which can limit the accuracy of flow and regurgitant fraction measurements. Goal(s): Our goal was to improve the SNR of phase contrast images at 0.55T using a bSSFP readout and maintain accurate flow quantification. Approach: We developed a through-plane phase contrast sequence with bSSFP readout and tested it in phantoms and healthy volunteers. Results: Significantly improved SNR (25.5±9.6 vs 8.2±2.9, p<0.001) and accuracy of velocities measurement were demonstrated in phantom. Accurate flow and constant high SNR were measured using our method in 8 subjects, for both breath-hold and free-breathing cases. Impact: Our proposed phase contrast with bSSFP readout can reliably evaluate aortic flow at 0.55T with higher SNR and VNR leading to more accurate quantification compared to conventional GRE based phase contrast method.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Good health and well-being
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies