Trade-off between Readout Duration and Concomitant Field Effect in 3D MRF at 0.55T
Motivation: The use of long readouts in 3D MRF improves SNR efficiency but also suffers from stronger concomitant field effects. Goal(s): To evaluate the trade-off between readout duration and concomitant field effects in 0.55T 3D MRF and evaluate mitigation using the MaxGIRF framework. Approach: We performed 0.55T 3D MRF scans of 14 readout durations on the NIST/ISMRM phantom and compared reconstructions with and without concomitant field effects compensation. We analyzed T1 array's standard deviations as a function of readout duration. Results: With correction, MRF T1 standard deviation reduced to stability due to improved SNR efficiency of long readouts, however, broke due to uncorrectable blurring. Impact: Practical trade-off between a readout duration and concomitant field effects are demonstrated. 0.55T 3D MRF benefits from a longer readout duration (6.1-14.7ms) provided a concomitant field effect correction.
Authors
- Krishna S. Nayak (ORCID: https://orcid.org/0000-0001-5735-3550)
- Zhibo Zhu (ORCID: https://orcid.org/0000-0002-2130-1158)
- Nam Lee (ORCID: https://orcid.org/0000-0001-5320-1672)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3574
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00