Influence of Body Coil Drive Mode on the MR Safety of Interventional Devices

Motivation: RF-induced heating of interventional devices is reduced at lower magnetic fields, but dangerous heating conditions can still occur during MR-guided interventions. Goal(s): To investigate the effect of body coil drive mode on RF-induced heating of interventional devices at 0.55T and 1.5T. Approach: Numerical simulations, E-field, transfer function and temperature measurements are used to evaluate the dependency of RF-induced heating of commonly used interventional devices on the phase and amplitude settings of body coil ports. Results: RF-induced heating can be reduced using individually adapted body coil drive modes. Impact: Dual-drive body coil technology can be used to generate implant-friendly electromagnetic fields, which can enable safe use of clinical intravascular devices during MR-guided interventions at 0.55T and 1.5T.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/3723
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Body Coil Drive Mode on the MR Safety of Interventional Devices

Oktay Algın, Simon Reiss, Michael Bock, Ali Çağlar Özen
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Influence of Body Coil Drive Mode on the MR Safety of Interventional Devices

Oktay Algın, Simon Reiss, Michael Bock, Ali Çağlar Özen
article en

Abstract

Motivation: RF-induced heating of interventional devices is reduced at lower magnetic fields, but dangerous heating conditions can still occur during MR-guided interventions. Goal(s): To investigate the effect of body coil drive mode on RF-induced heating of interventional devices at 0.55T and 1.5T. Approach: Numerical simulations, E-field, transfer function and temperature measurements are used to evaluate the dependency of RF-induced heating of commonly used interventional devices on the phase and amplitude settings of body coil ports. Results: RF-induced heating can be reduced using individually adapted body coil drive modes. Impact: Dual-drive body coil technology can be used to generate implant-friendly electromagnetic fields, which can enable safe use of clinical intravascular devices during MR-guided interventions at 0.55T and 1.5T.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Affordable and clean energy
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Influence of Body Coil Drive Mode on the MR Safety of Interventional Devices — Oktay Algın, Simon Reiss, et al. · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2024) | TGRS Research Map | TGRS