Feasibility of motional eddy current-reduced, passive eddy current shielding of MRI gradient coils
Motivation: High-performance gradient coils can induce substantial eddy current in the magnet causing heating and artifacts. Motional eddy current at high B0 exacerbates the problem. Goal(s): To investigate the feasibility of patterned, additional passive shielding to reduce both electromagnetically-indued and motional eddy currents. Approach: A proof-of-concept experiment was conducted in flat geometry, where shielding efficiency at 3T was compared between a solid copper and a patterned copper plate that mimicked the eddy current image of a driving coil. Results: The leakage field of patterned copper agreed with theory and was more localized in frequency and space. The shielding efficiency was improved at high frequencies. Impact: We demonstrated that vibration-induced motional eddy current of a passive conductive shield can be dramatically altered by eddy current cut-outs. The results can be used to build a compact, high-field scanners with reduced magnet-gradient interaction.
Authors
- Seung‐Kyun Lee (ORCID: https://orcid.org/0000-0001-7625-3141)
- Yihe Hua (ORCID: https://orcid.org/0009-0004-6674-9716)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3930
- Primary Topic
- Non-Destructive Testing Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00