Enhanced approach for three-dimensional myelin-weighted imaging using a novel flow suppression technique in the ViSTa
Motivation: ViSTa offers improved myelin-weighted image. However, the presence of flow artifacts was seen when scanning a subject with a high flow speed. Goal(s): This study aimed to construct a novel 3D ViSTa sequence by implementing a series of flow saturation pulses. Approach: Three flow saturation pulses were placed as evenly as possible between the 1st inversion pulse and the excitation pulse to effectively mitigate the presence of flow with various speeds. Results: The results from the proposed ViSTa sequence reveal a significant reduction in flow artifacts. When we conduct MS patient scans, lesions have a significantly diminished signal level, providing a distinct demarcation. Impact: The proposed method provides improved whole-brain covered 3D myelin-weighted images in a clinically reasonable scan time (< 7 min). Moreover, it shows good sensitivity to MS lesions. These features make the proposed method appealing for clinical neuroimaging applications.
Authors
- Mark J. Lowe (ORCID: https://orcid.org/0000-0002-7022-8436)
- Soleakhena Ken (ORCID: https://orcid.org/0000-0001-7904-5882)
- Sehong Oh (ORCID: https://orcid.org/0000-0003-2935-6580)
- Gawon Lee (ORCID: https://orcid.org/0000-0002-9955-1974)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2978
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00